ture during sudden large temperature changes. However, the threshold should
not be too small or the reset may be triggered by ordinary noise making the fil-
ter ineffective. The message “FILTER RESET” appears below the temperature
display whenever the filter resets to help alert you if there is a problem. The
threshold should be set so it is just larger than the maximum deviation caused
by noise. For many applications a threshold value of 0.01 to 0.1 is ideal. In
cases where the temperature variations are greater, the threshold value should
be increased. For temperature units (°C, K, °F) the threshold value is the devia-
tion in temperature that will cause reset. For resistance measurements the
threshold is given as a percentage of the measured resistance or reference resis-
tance, whichever is greater. For ratio measurements the threshold is given as a
percentage of the measured ratio or one, whichever is greater.
NOTE:
The filter threshold value must be greater than the noise in the
measurement or the filter will be ineffective.
The digital filter can also be reset manually at any time by pressing the DIGI-
TAL FILTER soft-key and then pressing Exit. The filter resets whenever the fil-
ter window is exited. The filter will also be reset if the system units are changed
unless the units are only changed from one type of temperature units to another,
e.g., °C to °F.
7.2.5
Reset Statistics
Pressing the RESET STATS soft-key causes the statistical registers to be
cleared. Any subsequent statistical calculations comprise only measurements
obtained from that time forward. Separate statistical registers and calculations
are used for each channel. The reset function resets the statistics of both chan-
nels simultaneously.
A variety of statistical functions can be applied to the measurements and the re-
sults displayed in various fields on the display. The available statistical func-
tions are maximum, minimum, spread, average, and standard deviation. (For
more information on setting up the display to show statistical calculations, see
Section 7.5).
7.3
Memory Menu
The MEMORY menu allows you to store measurements or arbitrary values in
memory. The numbers can then be displayed on the screen or used as refer-
ences for difference measurements (refer to Section 7.5.3). There are 100 avail-
able memory locations.
43
7 Front Panel Operation
Memory Menu
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......