7.6.4.4.1
Parameter
The PARAMETERS soft-key allows you to specify the channel and the type of
measurement to output. When you press the soft-key, a window appears allow-
ing you to specify the channel. If you select 1, the output will show only the
most recent measurement on Input 1. Likewise, 2 will show Input 2. If you se-
lect DEF (default) then the most recent measurement will be output regardless
of which channel was measured.
Next, a window will appear requesting you to specify a parameter to output.
For a complete description of the options, refer to Section 7.5.3.
7.6.4.4.2
Resolution
The RESOL soft-key allows you to set the resolution of the analog output. The
resolution is set in units per volt where the units depend on the specified pa-
rameter and system units. For example, if the resolution is 100, the parameter is
T (temperature in the current units), and the system units are set for °C, for ev-
ery 100°C increase in the measurement, the analog output will increase 1 volt.
The smaller the resolution, the smaller the range of measurements that can be
output.
7.6.4.4.3
Offset
The OFFSET soft-key allows you to set the offset of the analog output. The off-
set is the measurement value, in the appropriate units, which will output 0V.
For example, if the offset is set to 25, the output will be 0V when 25°C is mea-
sured. The following formulas show the relationship between the output volt-
age and the measurement and how RESOL and OFFSET affect it.
Output volts
Measurement OFFSET
RESOL
[
]
(
)
=
−
Measurement
RESOLxOutput volts
OFFSET
=
+
[
]
7.6.4.4.4
Center
The CENTER button is used to quickly set the offset value to the most recent
measurement value. The display does not change.
7.6.4.4.5
Cal
The accuracy of the analog output can be improved by calibration. This can be
1575/1590 Thermometer Readout
User’s Guide
92
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......