Innovative High Temperature Fiber Optic Infrared Pyrometers
Installation
25
Figure 11: Caution / Attention hint for devices with ordered LASER sighting option
Ensure fiber optic cable is connected before turning on the laser!
5.3. Aiming and Focusing
An effective aiming technique is to adjust the sensor head until the highest reading is observed
on the internal display. When the highest reading is reached, hold the unit in place and secure
the mounting base.
Make sure that the sensor is in 1-color mode when using this aiming technique!
Another aiming can be done by means of a battery powered aiming light. Simply loosen the
compression sleeve holding the fiber optic cable, loosen the screw at the heater block, and pull
the cable out of the heater block approximately 7 mm (0.25 in). Raise the fiber optic cable
enough to slip the aiming light onto the end. Align the light beam on the target.
Содержание endurance series
Страница 4: ......
Страница 8: ...iv...
Страница 10: ...vi...
Страница 24: ...Endurance Series Users Manual 12 3 4 1 1 Measurement Spot Size Diagrams for F0 Narrow Focus Option...
Страница 25: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 13...
Страница 26: ...Endurance Series Users Manual 14 3 4 1 2 Measurement Spot Size Diagrams for F1 Close Focus Option...
Страница 27: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Technical Data 15...
Страница 28: ...Endurance Series Users Manual 16 3 4 1 3 Measurement Spot Size Diagrams for F2 Infinity Focus Option...
Страница 69: ...Innovative High Temperature Fiber Optic Infrared Pyrometers Signal Processing 57 defaults are restored...