Extend measurement range manually
Measurement range may be extended manually in both Active Clock and Static measurement
modes. This range extension allows the adjustment of the laser beam such that it does not
miss the detector surface when measuring shafts with gross misalignment or angular mis-
alignment over large distances. During measurement, manual extension is prompted by
accessing the XY view before 'Laser End' is displayed.
l
If the laser dot (
1
)on the display continues to move further away from the center of the
detector screen while rotating shafts to take measurements, tap the detector area (
2
) to
access the "XY view" screen.
l
After accessing the "XY view", use the yellow horizontal angle adjustment knob and the
vertical position adjustment thumbwheel to adjust the laser dot such that it is positioned
inside or very close to the square target.
Note
During the laser adjustment procedure, avoid readjusting the sensor.
SHAFTALIGN touch
65
On-board help
Содержание db PRUFTECHNIK SHAFTALIGN touch
Страница 1: ...SHAFTALIGN touch On board help...
Страница 2: ...SHAFTALIGN touch On board help Version 2 3 Edition 07 2020 Part No DOC 26 200 EN...
Страница 24: ...SHAFTALIGN touch 23 On board help...
Страница 42: ...After target values have been entered tap to proceed SHAFTALIGN touch 41 On board help...
Страница 63: ...62 Version 2 3 On board help...
Страница 89: ...Note If no template is selected all template list options are unavailable 88 Version 2 3 On board help...
Страница 127: ...W Wireless connection 21 X XY view 81 XY View 53 SHAFTALIGN touch 126 On board help...