990DSL
Users Guide
4-34
3.
With the strapping prompt screen displayed, press
K
.
This starts the second part of the K-Test, during which the tester
measures loop resistance and performs a nulling bridge operation on
the pair
with the far-end shorted.
After completion, the tester displays resistances to the fault and strap,
as in the normal RFL process.
4.
Press
A
(
Distance to Fault
) to view distances, as is normally done.
Load Coils Test
The Load Coils test performs a frequency sweep of the pair (either spare or
working POTS) to determine the number of load coils present, and the
approximate distance to the first coil.
Setting Up a Load Coils Test
The Load Coils test requires that the
predominant
wire-gauge to the first load
coil be entered in Setups, for the most accurate distance-to-first-load results.
Load Coils Test Results
To run the Load Coils test, select the test on the POTS Toolbox and press
K
.
Figure 4-21 shows the results of a Load Coils test that was conducted on a pair
with six loads:
Содержание 990DSL
Страница 4: ......
Страница 12: ...990DSL Users Manual viii...
Страница 26: ...990DSL Users Guide 1 12...
Страница 64: ...990DSL Users Guide 3 26...
Страница 138: ...990DSL Users Guide 4 74...
Страница 202: ...990DSL Users Guide 6 20...
Страница 220: ...990DSL Users Manual 8 6...
Страница 221: ...Appendices Appendix Title Page A Features and Specifications A 1 B Replacement Parts and Accessories B 1...
Страница 222: ...990DSL Users Guide...
Страница 232: ...990DSL Users Guide A 10...
Страница 238: ...990DSL Users Guide B 6...
Страница 242: ...990DSL Users Manual 4 Z Zero Leads utility 3 20...