POTS Testing and Fault Location
The POTS Toolbox
4
4-17
CPE Correction
If a bad Ground connection, open sheath, or far-end CPE termination is
present, the tester will measure a 3T Capacitance Ratio that is much lower than
normal. As a result, the 990 will display
Open Ground or CPE
on the results
screen and will also provide a
CPE Correction
softkey
C
. If a bad ground
connection or open sheath is not suspected, press the CPE Correction softkey.
The unit will then display corrected distances, subtracting CPE effects.
Adjusted Opens Test
A cable manufacturer’s data can vary by as much as 7% among cables.
Therefore, for greater accuracy, it is strongly recommended that you first
perform an Opens test on a
known good
pair in the cable. This data can then be
saved as a Custom Cable Type for further use on faulted pairs in the same
cable. Alternatively, if the length of a good pair is known by another more
accurate means, you can enter that length and store the data as a Custom Cable
Type. This is called an
Adjusted Opens
measurement.
Figure 4-8 shows the results from a balanced pair being used to create a
Custom Cable Type for future usage on other pairs in the same cable
To run an Adjusted Opens test, do the following:
1.
Run an Opens Test on a good, balanced pair in the cable.
2.
Press
A
(
Adjusted Opens
).
A screen like the one in Figure 4-8 is displayed:
Содержание 990DSL
Страница 4: ......
Страница 12: ...990DSL Users Manual viii...
Страница 26: ...990DSL Users Guide 1 12...
Страница 64: ...990DSL Users Guide 3 26...
Страница 138: ...990DSL Users Guide 4 74...
Страница 202: ...990DSL Users Guide 6 20...
Страница 220: ...990DSL Users Manual 8 6...
Страница 221: ...Appendices Appendix Title Page A Features and Specifications A 1 B Replacement Parts and Accessories B 1...
Страница 222: ...990DSL Users Guide...
Страница 232: ...990DSL Users Guide A 10...
Страница 238: ...990DSL Users Guide B 6...
Страница 242: ...990DSL Users Manual 4 Z Zero Leads utility 3 20...