Fluke 983 Скачать руководство пользователя страница 1

®

 

Fluke 983 

Particle Counter 

 

 

 

Users Manual 

  

 
 
 
PN 2414721 
January 2005  Rev. 2, 1/07 

© 2005-2007 Fluke Corporation. All rights reserved. Printed in U.S.A. 
All product names are trademarks of their respective companies.

 

Содержание 983

Страница 1: ...Fluke 983 Particle Counter Users Manual PN 2414721 January 2005 Rev 2 1 07 2005 2007 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their respective companies ...

Страница 2: ...ice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepaid FOB Destination If Fluke deter mines that failure was caused by neglect misuse contamin...

Страница 3: ...acting Fluke 1 Safety Information 2 Compliance 4 Quick Start 4 Shipping Contents 5 The Keypad 6 The Backlight 6 The Base 7 The Display 7 The Main Screen 8 The Setup Screens 9 The Clock Setup Menu 10 Label Menu 10 Reviewing the Buffer 11 Using the Particle Counter 12 ...

Страница 4: ...Setting a Delay Between Samples 17 Setting Temperature Mode 18 Restoring Default Settings 18 Setting the I O Port to Download Data to a PC 18 Transferring Data to a Personal Computer 18 Control 19 Labels 20 Time 20 Printing the Samples to a Serial Printer 20 Connecting the Printer to the Particle Counter 20 To Print a Single Sample 20 To Print a Single Sample Stored in the Buffer 21 To Print all o...

Страница 5: ...Contents continued iii Purging the Particle Counter Sensor 23 Troubleshooting 24 Specifications 26 Parts and Accessories 27 ...

Страница 6: ...Fluke 983 Users Manual iv ...

Страница 7: ...v List of Tables Table Title Page 1 Symbols 2 2 Troubleshooting 25 3 Replaceable Parts and Accessories 29 ...

Страница 8: ...Fluke 983 Users Manual vi ...

Страница 9: ...pping Contents 5 2 Fluke 983 Keypad 6 3 The Base of the Fluke 983 7 4 Main Screen 8 5 Setup Screen 1 9 6 Setup Screen 2 9 7 Clock Setup Menu 10 8 Label Menu 12 9 Using the High Purity Tubing and Barbed Isokinetic Probe 14 10 Fluke 983 Replaceable Parts 28 ...

Страница 10: ...Fluke 983 Users Manual viii ...

Страница 11: ... cable and FlukeView Forms Documenting Software provided with the unit Applications The Particle Counter can be used for Monitoring cleanrooms Indoor air quality investigations Monitoring gowning rooms Testing filter seals Locating particle contamination sources Monitoring particle size distributions Contacting Fluke To contact Fluke or for service call one of the following telephone numbers USA 1...

Страница 12: ...the entire Users Manual before using the Particle Counter Safety symbols used in this manual and on the Particle Counter are displayed in Table 1 Table 1 Symbols Symbol Meaning W Risk of Danger Important information Refer to manual X Hazardous voltage Risk of electrical shock Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information P Conforms to re...

Страница 13: ...ified service personnel Inspect the Particle Counter before use Do not use it if it appears damaged Always use the ac adapter charger and connector supplied with the Particle Counter appropriate for the voltage and outlet of the country or location in which you are working WCaution To avoid possible damage to the Particle Counter Avoid using the Particle Counter in an excessively dirty or dusty at...

Страница 14: ...2001 Federal Standard 209E Quick Start The Particle Counter may be used directly out of the box To immediately use the Particle Counter 1 Read Safety Information 2 Remove the Isokinetic Probe Cover the black cap 3 Press the POWER keypad 4 Press 1 to start sampling Be sure to read the entire Users Manual for more detailed information ...

Страница 15: ...sh B Isokinetic Probe and Cap J Safety Information Multi language C AC Adapter Charger K Intake Nozzle and Cap D RS 232 Modular Cable L Power Cord E in High Purity Tubing 2 ft M Hardshell Carrying Case F RJ 45 to DB 9 Connector N Zero Count Filter Adapter G Zero Count Filter O Isokinetic Probe with Barb H 983 Users Manual CD P FlukeView Forms Documenting Software CD Figure 1 Fluke 983 Shipping Con...

Страница 16: ... backlight turns off to conserve battery power If the backlight goes out press any function or arrow key to turn it back on 983 bak01f eps Power Key Turns the Particle Counter on and off Arrow Keys Press to navigate up and down through the different sections of the menus or to control the display contrast when on the Main Screen F1 F2 F2 F4 Function Keys Each function key corresponds to a name abo...

Страница 17: ... Power Tripod Mount I O Port bak08f eps Figure 3 The Base of the Fluke 983 The Display The LCD Display shows several different screens and menus Navigate through each using the keypad The following sections explain the different screens used by the Particle Counter The following screens appear on the display Main Screen Setup Screen pages 1 and 2 Clock Setup Menu Label Setup Menu The bottom of the...

Страница 18: ...channels Particle counts Relative humidity RH Air sample temperature C or F Sample location name Battery charge indicator charge remaining Time indicator represents time to complete Sample volume Particle count mode Total counts volume The parameters on the Main Screen are selected by using the Setup Screens the Clock Setup Menu and the Label Setup Menu F1 F2 F3 F4 bak03f eps Figure 4 Main Screen ...

Страница 19: ...tup Screen 1 press 1 PAGE 2 see Figure 6 2 Use and to move up and through each parameter Use 2 left arrow and 3 right arrow to navigate within each screen entry that is highlighted See Figures 5 and 6 3 Press 4 RETURN to store the setup changes and to return to the Main Screen F1 F2 F3 F4 Password Disabled bak02f eps Figure 5 Setup Screen 1 F1 F2 F3 F4 bak06f eps Figure 6 Setup Screen 2 ...

Страница 20: ...change the format of the date from dd mm yyyy to mm dd yyyy 1 Move the cursor down to Date 2 Select 1 PROGRAM 3 Select 4 FORMAT The date will switch to the next version 4 Press 1 ENTER to store the choice To use a 24 hour clock or a 12 hour clock 1 Move the cursor down to Time 2 Select 1 PROGRAM 3 Select 4 FORMAT The clock will switch to a 24 or 12 hour version 4 Press 1 ENTER to store the choice ...

Страница 21: ...u without storing the label name 7 Continue to edit label names by following the previous process When finished press 4 RETURN to return to the Main Screen Notice that the Main screen now has the sample label in the upper right hand corner You may also use the FlukeView Forms Documenting Software provided to change the label names Refer to Transferring Data to a Personal Computer Reviewing the Buf...

Страница 22: ...the Isokinetic Probe In a clean room application the zero count filter verifies that the counter is not counting electrical noise signals from internal components and is not subject to external interference In other applications the zero count filter purges the sensor immediately after a high concentration sampling If the unit fails see Purging the Particle Counter Sensor Set Up To set up the Part...

Страница 23: ...ew the threaded Isokinetic Probe 2 Replace it with the intake nozzle 3 Connect the high purity tubing to the intake nozzle 4 Connect the Barbed Isokinetic Probe to the open end of the tubing Intake Nozzle Barbed Isokinetic Probe High Purity Tubing bak10f eps Figure 9 Using the High Purity Tubing and Barbed Isokinetic Probe ...

Страница 24: ...eeded see Setting the Alarm For example if the limit is set to 10 the counter beeps when the count first reaches 10 and beeps again for each multiple of 10 In Audio Mode the display shows counts as in Totalize Mode 1 On Page 1 of the Setup Screen use and to highlight Cnt Mode If coming from the Main Screen Cnt Mode should already be highlighted 2 Use 2 left arrow and 3 right arrow to select other ...

Страница 25: ...ither liters or cubic feet based on the time selected Setting the Method of Counting Data The choices for data counting are Cumulative Cumulative counting mode includes all particles that are greater than or equal to the particle size selected in the Sample Volume field Differential Differential counting mode includes all particles that are greater than or equal to the particle size selected in th...

Страница 26: ...cle Counter s serial number in the upper right corner of the display 5 Press 1 CHANGE 6 Use and to change the selected character of the password 7 Press 2 left arrow or 3 right arrow to move the cursor to the next character 8 Continue to do this until you have entered the desired password 9 When finished press 4 ENTER to save the password or press 4 ESCAPE to exit the password menu without saving ...

Страница 27: ...arrow or 3 right arrow to move the alarm limits from zero no alarm to 100 000 in factors of 10 4 Once the alarm limit is selected press to select Alarm Size 5 Press 2 left arrow and 3 right arrow to display the various particle sizes 6 Select the desired particle size and press 4 RETURN to return to the Main Screen Note When the alarm sounds pressing 1 STOP once will mute the alarm Pressing 1 STOP...

Страница 28: ...ss or to move the cursor to I O Port 2 Use 2 left arrow and 3 right arrow to select Serial 9600 3 Press 4 RETURN to return to the Main Screen Note Serial 9600 is the default I O Port setting for software version 1 1 and cannot be changed Transferring Data to a Personal Computer Data that is collected and stored in the Particle Counter s memory can be downloaded to a personal computer PC using the ...

Страница 29: ...r in the Flags column of the Virtual meter individually or in combination 0 Normal 1 Sensor Error Samples are invalid indicates a calibration failure or sensor failure condition 2 Count Alarm The user defined alarm limit has been exceeded 4 Low Battery The battery charge was low during the sample Recharge the battery as soon as possible 8 Counter Error Samples are invalid This flag indicates a har...

Страница 30: ...Fluke for instructions on how to update your software Fluke offers a complete Thermal Printer Kit to take advantage of the printing capabilities of the Particle Counter The Fluke 983 Thermal Printer kit includes Seiko DPU 414 Thermal Printer Seiko PW 4007 U1 120 V ac adapter or PW 4007 E1 220 V ac adapter Seiko BP 4005 battery pack 1 roll thermal paper RJ 45 to DB 9 serial printer adapter To use t...

Страница 31: ... Counter stores each measurement in a 5000 record buffer Data is stored in a first in first out order When the 5001st record is saved the first record is deleted leaving 5000 records total Clearing the Buffer At some point the user may wish to clear the Particle Counter s buffer To do this 1 Go to Setup Screen 2 and press or to move the cursor to Buffer Count Buffer Count displays the number of sa...

Страница 32: ...e battery is partially charged The ac adapter is used to recharge the battery as well as power the Particle Counter This feature allows the counter to be used while the battery is charging To connect the ac adapter charger 1 Attach the power cord to the end of the charger 2 Plug the charger into the 12V jack on the base of the Particle Counter 3 Plug the ac adapter into an ac outlet The Particle C...

Страница 33: ... adapter onto the end of the Zero Count Filter Note the arrow on the filter denoting airflow direction 2 Connect the adapter directly onto the end of the Isokinetic Probe 3 Operate the unit for approximately 15 minutes 4 On the average not more than one particle greater than 0 3 µm in five minutes or not more than one particle per 0 5 cubic foot should be apparent This purges the Particle Counter ...

Страница 34: ...he Particle Counter does not turn on contact Fluke for assistance LOW BATTERY displays Battery voltage is low Refer to Charging the Battery and Using the AC Adapter Particle Counter battery does not charge A problem with the battery is indicated Disconnect the Particle Counter from the ac adapter charger Check the line cord If the problem persists contact Fluke After fully charging the battery in ...

Страница 35: ... shows CAL ALARM 1 The sample exceeded the concentration limit too many particles which produced a temporary alarm No corrective measure is required other than to stop sampling in that environment 2 The sensor is contaminated Refer to Purging the Particle Counter Sensor 3 The laser has dropped below acceptable levels If the problem persists contact Fluke No temperature humidity reading or an incor...

Страница 36: ...e time counts relative humidity temperature sample volumes alarms label Display Graphic liquid crystal with backlight Count Modes Concentration totalize audio Delay Time 0 to 24 hours Sample Inlet Isokinetic probe Interface RS 232 and RS 485 via RJ 45 Vacuum Source Internal pump flow controlled Dimensions 4 5 x 8 25 x 2 25 11 43 x 20 96 x 5 72 cm Weight 2 2 pounds 1 0 kg Environmental Operating 10...

Страница 37: ...Particle Counter Parts and Accessories 27 Parts and Accessories To order parts and accessories for the Particle Counter refer to Table 3 and Figure 10 bak09f eps Figure 10 Fluke 983 Replaceable Parts ...

Страница 38: ...Modular Cable F 2421455 High Purity Tubing 2 ft G 2421392 RJ 45 to DB 9 Connector H 2421428 Zero Count Filter I 2414717 983 Users Manual CD J 2414721 Users Manual English K 2414700 Safety Information Multi language L 2421443 Intake Nozzle and Cap M 2421361 US Power Cord N 2414693 Hardshell Carrying Case O 2414668 Zero Count Filter Adapter P 2424376 Barbed Isokinetic Probe Q 1578384 FlukeView Forms...

Страница 39: ...wn 2543469 983 US Printer Kit 120V version Seiko DPU 414 Thermal Printer Seiko PW 4007 U1 120 V AC Adapter Seiko BP 4005 Battery Pack 1 Roll Thermal Paper RJ 45 to DB 9 serial Printer Adapter Not Shown 2547127 983 Euro Printer Kit 220V version Seiko DPU 414 Thermal Printer Seiko PW 4007 E1 220 V AC Adapter Seiko BP 4005 Battery Pack 1 Roll Thermal Paper RJ 45 to DB 9 serial Printer Adapter ...

Страница 40: ...Fluke 983 Users Manual 30 ...

Отзывы: