975CK English Instruction Sheet 

Page  

 

 

®

 

975CK 

Gas Sensor Calibration Kit 

Instruction Sheet 

Calibration 

Calibration due dates are tracked using the Meter’s clock and 
stored in non-volatile memory. Due dates can be configured by 
the user from one to 365 days. When the Meter reaches its 
calibration due date, it alerts the user but will continue to operate. 

The Meter’s CO and CO

2

 sensors can be calibrated by the user 

or returned to Fluke for service. See “Contacting Fluke”. The 
recommended calibration interval is 1 month for CO and 1 year 
for CO

2

Gas canisters are available as an accessory allowing the user to 
calibrate the CO and CO

2

 sensors. 

Install the calibration cap and attach the hose to the correct gas 
canister; mixed CO and CO

2

 for calibration gas, Nitrogen for 

zeroing gas when calibrating CO

2

. See Figure 1 

Note 

Calibration gas is applied at the rate of ½ liter/minute for 
2 minutes. 

F1

F1

VELOCITY

PROBE

F1

 

eha001.eps 

Calibration Connections 

 
 
PN 

2768489

 

October 2006 

© 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. 
All product names are trademarks of their respective companies. 