Equipment required for this test:
•
Measuring Receiver
•
Attenuator, 6 dB, Type-N male / female, 50
Ω
•
50
Ω
Leveling Head (supplied with UUT)
Note
In the following tests, the tolerances shown refer to specifications listed in
Chapter 1 of this manual. It may be necessary, in some cases, to alter the
test limits based on the uncertainty of the actual equipment used. For
example, if the Instrument specification is ±0.025 dB and the measuring
receiver uncertainty is ± (0.015 dB + 0.005 dB per 10 dB step) the test limit
for 30 dB would be ±0.039 dB (the root-sum-square of 0.025 and 0.030).
Use the following procedure to verify the attenuation accuracy of the UUT output relative
to +16 dBm, 50
Ω
:
1.
Connect the Leveling Head to the RF INPUT of the measuring receiver via a 6 dB
attenuator and connect the REF FREQUENCY OUTPUT (rear panel) from the UUT
to the spectrum analyzer Ext Ref In (on rear). See Figure 5-7 for equipment
connections.
9610A/AF REFERENCE SOURCE
10V
PK
MAX
RF
COMMON
96xx
HEAD INTERFACE
RF OUTPUT
CONTROL
CONNECT
96xx HEAD
ONLY
STBY
NEXT
CHAR
6
9
5
8
4
7
7
3
2
1
-
-
0
.
SPACE
EXP
BKSP
ALPHA
LOCK
LOCK
DEF
MNO
ABC
JKL
/
/
GHI
WXYZ
TUV
PQRS
ENTER
SINE
MOD
SWEEP
UNITS
SETUP
OPER
CAL
IEEE-488
SH1 AH1 T6 L4 SR1
RL1 PP0 DC1 DT0 C0 E2
MADE IN UK
FLUKE CORPORA
TION
www.fluke.com
TO BE
APPLIED FOR
CONFORMS
TO UL STD. UL61010A-1.
CERTIFIED TO CAN/CSA STD. C22.2 No.1010.1-92
SAFETY W
ARNING
NOTE
DISCONNECT PO
WER AND SIGNAL LEADS
BEFORE REMO
VING CO
VERS
.
FOR CONTINUED PR
OTECTION
AGAINST ELECTRIC SHOCK
THE PO
WER CORD PR
OTECTIVE CONDUCT
OR MUS
T BE
CONNECTED
TO SAFETY EAR
TH ( GR
OUND ).
FOR CONTINUED PR
OTECTION
AGAINST FIRE
USE 250V FUSE OF CORRECT RA
TING.
READ OPERA
TORS MANU
AL BEFORE USE.
NO USER SER
VICEABLE P
ARTS CONT
AINED, DO NOT
REMOVE CO
VERS
, HAZARDOUS
VOLTAGES PRESENT
.
REFER SER
VICE TO QUALIFIED PERSONNEL.
115
230
T4.0 AH 250V
T2.0 AH 250V
LINE VOLTAGE
SELECTION
FUSE 5 x 20mm
IEC 127
POWER INPUT
230 VA MAX. FREQ 47 - 63 Hz
SERIAL POR
T
FACTORY SET
REF FREQUENC
Y
INPUT
REF F REQUENC
Y
OUTPUT
6 7 8
ENABLE
DISABLE
1V PK INTO
TTL INTO
SWEEP TRIGGER
INPUT :
5V PK
AUX PROTECTIVE
EARTH (GROUND)
RF COMMON
RF COMMON
5V PK
MODULATION
LEVELLING AND
FREQUENCY PULL
INPUT
+
_
50
5V PK
+
_
50
+
OUTPUT : TTL NOM
INTO 1k
+
1k NOM
NOM
50
10V PK MAX
Front
Rear
50
Ω
Leveling Head
6 dB Attenuator
RF Input
Ext Ref Input
Spectrum Analyzer
Ref Frequency
Output
Instrument
ead220f.eps
Figure 5-7. Equipment Connections - Attenuation Accuracy Test (50
Ω
)
Содержание 9640A Series
Страница 8: ......
Страница 16: ...9640A Instruction Manual viii...
Страница 22: ...9640A Instruction Manual 1 2...
Страница 40: ...9640A Instruction Manual 2 2...
Страница 108: ...9640A Instruction Manual 4A 2...
Страница 138: ...9640A Instruction Manual 4C 2...
Страница 164: ...9640A Instruction Manual 4D 2...
Страница 168: ...9640A Instruction Manual 4D 6...
Страница 170: ...9640A Instruction Manual 4E 2...
Страница 174: ...9640A Instruction Manual 4E 6...
Страница 176: ...9640A Instruction Manual 4F 2...
Страница 184: ...9640A Instruction Manual 4F 10...
Страница 186: ...9640A Instruction Manual 4G 2...
Страница 190: ...9640A Instruction Manual 4G 6...
Страница 192: ...9640A Instruction Manual 5 2...
Страница 272: ...9640A Instruction Manual 5 82...
Страница 274: ...9640A Instruction Manual 6 2...
Страница 284: ...9640A Instruction Manual 7 2...
Страница 309: ...8 1 Chapter 8 Lists of Replaceable Parts Title Page Introduction 8 3 How to Obtain Parts 8 3 Parts Lists 8 4...
Страница 310: ...9640A Instruction Manual 8 2...
Страница 316: ...9640A Instruction Manual 8 8...
Страница 320: ...9640A Instruction Manual A 4...