*CLS
The *CLS common command clears the status data structures by clearing all
event registers and the error queue. It does not clear enable registers and
transition filters. It clears any pending *WAI, *OPC, and *OPC?.
*ESE NR1
Sets the enable bits of the standard event enable register. This enable register
contains a mask value for the bits to be enabled in the standard event status
register. A bit that is set true in the enable register enables the corresponding
bit in the status register. An enabled bit will set the ESB (Event Status Bit) in the
Status Byte Register if the enabled event occurs.
*ESR?
Reads out the contents of the standard event status register. Reading the
Standard Event Status Register clears the register.
*IDN?
Reads out the manufacturer, model, serial number, Firmware level for main and
GPIB program in an ASCII response data element.
Response is <Manufacturer> , <Model> , <Serial Number>, <Firmware Level>.
Eg: Fluke,9640A,123456,1.34 or Fluke,9640A-LPN,54321,2.00
*OPC
The Operation Complete command causes the device to set the operation
complete bit in the Standard Event Status Register when all pending selected
device operations have been finished. (Currently only AM settings have
significant delays.)
*OPC?
Operation Complete query. The Operation Complete query places an ASCii
character 1 into the device’s Output Queue when all pending selected device
operations have been finished. (Currently only AM settings have significant
delays.)
*OPT?
Response is a list of all detectable options present in the instrument, with
absent options represented with an ASCii ‘0’.
The first item in the list represents the 8662/3 emulation option.
*PSC NR1
Enables/disables automatic power-on clearing. The status registers listed below
are cleared when the power-on status clear flag is 1. Power-on does not affect
the registers when the flag is 0.
*RST
The Reset command resets the instrument. It is the third level of reset in a 3-
level reset strategy, and it primarily affects the instruments functions, not the
IEEE 488 bus.
*SRE NR1
The Service Request Enable command sets the service request enable register
bits. This enable register contains a mask value for the bits to be enabled in the
status byte register. A bit that is set true in the enable register enables the
corresponding bit in the status byte register to generate a Service Request.
*STB?
Reads out the value of the Status Byte. Bit 6 reports the Master Summary
Status bit (MSS), not the Request Service (RQS). The MSS is set if the
instrument has one or more reasons for requesting service.
*TST?
The self-test query causes an internal self-test and generates a response
indicating whether or not the device completed the self-test without any
detected errors.
*WAI
The Wait-to-Continue command prevents the device from executing any further
commands or queries until execution of all previous commands or queries have
been completed.
Содержание 9640A Series
Страница 8: ......
Страница 16: ...9640A Instruction Manual viii...
Страница 22: ...9640A Instruction Manual 1 2...
Страница 40: ...9640A Instruction Manual 2 2...
Страница 108: ...9640A Instruction Manual 4A 2...
Страница 138: ...9640A Instruction Manual 4C 2...
Страница 164: ...9640A Instruction Manual 4D 2...
Страница 168: ...9640A Instruction Manual 4D 6...
Страница 170: ...9640A Instruction Manual 4E 2...
Страница 174: ...9640A Instruction Manual 4E 6...
Страница 176: ...9640A Instruction Manual 4F 2...
Страница 184: ...9640A Instruction Manual 4F 10...
Страница 186: ...9640A Instruction Manual 4G 2...
Страница 190: ...9640A Instruction Manual 4G 6...
Страница 192: ...9640A Instruction Manual 5 2...
Страница 272: ...9640A Instruction Manual 5 82...
Страница 274: ...9640A Instruction Manual 6 2...
Страница 284: ...9640A Instruction Manual 7 2...
Страница 309: ...8 1 Chapter 8 Lists of Replaceable Parts Title Page Introduction 8 3 How to Obtain Parts 8 3 Parts Lists 8 4...
Страница 310: ...9640A Instruction Manual 8 2...
Страница 316: ...9640A Instruction Manual 8 8...
Страница 320: ...9640A Instruction Manual A 4...