Dry-well Calibrator
Operation
17
High Limit
The High Limit parameter adjusts the upper set-point temperature. The factory default and maximum
are set to 125 °C (257 °F). The minimum setting is 50 °C (122 °F). For safety, a user can adjust the
High Limit down so the maximum temperature set-point is restricted.
HL
Flashes HL and then displays the setting
125
Flashes the current value and then displays the value for adjustment
125
Current High Limit setting
Press
u
or
d
to adjust the setting.
100
New High Limit setting
To accept the new setting, press
SET
. Press
EXIT
to continue without storing the new value.
S
Accept the new High Limit setting
Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by:
SErL
Serial RS-232 interface parameters menu
Press
SET
to enter the menu. The serial interface parameters menu contains parameters which
determine the operation of the serial interface. The parameters in the menu are: baud rate, sample
period, duplex mode, and linefeed.
Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting determines the serial
communications transmission rate.
The baud rate parameter is indicated by::
bAUd
Flashes bAUd for one second and then displays the setting
2400b
Current baud rate
The BAUD rate of the serial communications may be programmed to 300, 600, 1200, 2400 (default),
4800, or 9600 BAUD. Use the up or down arrows to change the BAUD rate value.
4800 b
New baud rate
Press
SET
to accept the new setting or
EXIT
to abort the operation and skip to the next parameter in
the menu.