Dry-well Calibrator
Operation
13
Set-point Value
The set-point value may be adjusted after selecting the set-point memory and pressing
SET
.
4 50.
Set-point 4 value in °C
If the set-point value does not need to be changed, press and hold
EXIT
to resume displaying the well
temperature. To change the set-point value, press
SET
and then press
u
or
d
.
60.
New set-point value
When the desired set-point value is reached, press
SET
to accept the new value and access the
temperature scale units selection. If
EXIT
is pressed, any changes made to the set-point are ignored.
S
Accept new set-point value
To change temperature units, see
.
Scan
The scan rate can be set and enabled so that when the set-point is changed the dry-well heats or cools
at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the
dry-well heats or cools at the maximum possible rate.
Scan Control
The scan is controlled with the scan on/off function that appears in the main menu after the
temperature scale units.
Sc
Sc flashes for one second and then the current scan setting displays
Off
Scan function off
Press
u
or
d
to toggle the scan on or off.
On
Scan function on
Press
SET
to accept the present setting and continue. Press
EXIT
to cancel.
S
Accept scan setting