User’s Guide
12
The proportional band width is set at the factory. The proportional band width is easily adjusted from
the front panel. Set the width to discrete values in degrees C or F
,
which depends on the selected units.
The proportional band adjustment is accessed within the secondary menu. Push
SET
and
EXIT
to
enter the secondary menu and show the heater power. Then push
SET
to access the proportional
band.
Push
SET
and
EXIT
Access heater power in secondary menu
SEC
Flashes for secondary menu, then shows the heater power
13.0 P
Heater power in percent
Push
SET
Access proportional band
ProP
Flashes ProP and the current setting
4.1
Proportional band setting
To change the proportional band
,
push
UP
or
DOWN
.
10.0
New proportional band setting
To accept and store the new setting push
SET
. Push
EXIT
to continue without storing the new value.
Controller Configuration
The controller has several configuration and operating options and calibration parameters which are
programmable through the front panel. To access these, push
SET
from the secondary menu after the
proportional band. The display shows Config and then PAr for the first of three sets of configuration
parameters- operating parameters, serial interface parameters, and calibration parameters. To select
the menus
,
push
UP
and
DOWN
and then push
SET
(see Figure
Operating Parameters
The operating parameters menu is indicated by,
PAr
Operating parameters menu
Push
SET
to enter the menu. The operating parameters menu contains the HL (High Limit) parameter
which adjusts the upper set-point temperature. The factory default and maximum are set to 350 C. For
safety, adjust the HL down to restrict the maximum temperature set-point.
HL
Flashes HL, then shows the current value
H=350
Current HL setting
Adjust the HL parameter with
UP
or
DOWN
.
H=300
New HL setting
Push
SET
to accept the new temperature limit.