True-rms Multimeter
Calibration Adjustment
21
If the Meter rotary knob is not in the correct position, or if the measured
value is not within the anticipated range of the input value, the Meter emits
a double beep and will not continue to the next step.
Some adjustment steps take longer to execute than others (10 to 15
seconds). For these steps, the Meter will beep when the step is complete.
Not all steps have this feature.
8.
After the final step, the display shows "End" to indicate that the calibration
adjustment is complete. Press
D
to go to meter mode.
Notes
Set the calibrator to Standby prior to changing the function switch position
and or after completing adjustment of each function.
If the calibration adjustment procedure is not completed correctly, the
Meter will not operate correctly.
Table 8. Calibration Adjustment Steps
Function
(Switch Position)
Input Terminal
Adjustment Step
Input Value
C-01
600.0 mV, 60 Hz
C-02
600.0 mV, 20 kHz
C-03
6.000 V, 60 Hz
C-04
6.000 V, 20 kHz
C-05
60.00 V, 60 Hz
C-06
60.00 V, 20 kHz
C-07
600.0 V, 60 Hz
K
(AC Volts)
C-08
600.0 V, 10 kHz
C-09 6.000
V
C-10 60.00
V
L
(DC Volts)
C-11 600.0
V
C-12 600.0
mV
m
L
(DC Millivolts)
C-13 60.00
mV
C-14 600.0
C-15 6.000
k
C-16 60.00
k
C-17 600.0
k
C-18 6.000
M
C-19 0.000
e
(Ohms)
C-20 50.0
M
O
(Diode Test)
I
C-21 3.000
V
C-22
6.000 A, 60 Hz
A
(Amps)
A
C-23
6.000 A dc