background image

80T-150U-4011

Sensing Probe and Cable Assembly

White

Black

F4.eps

Figure 4. 80T-150U Temperature Probe Schematic

Replaceable Parts

5HI'HV

31

'HVFULSWLRQ

BT1

738179

80TK-8011         ,CONNECTOR, BATTERY

C1 C3-C7

106146

CAPACITOR,SMR,CAP,CER,0.01UF,+-10%,50V,X7R,0805

P1 P2

736033

80TK-8009,PLUG,BANANA

Q1 Q2

820910

PNP,MMBT5087,SMR,TRANSISTOR,SI,PNP,50V,225MW,SOT-23

R11

928903

RESISTOR,SMR,RES,CERM,464K,+-1%,0.1W,100PPM,0805

R12

928866

RESISTOR,SMR,RES,CERM,100K,+-1%,0.1W,100PPM,0805

R14

802733

RESISTOR,SMR,RES,MF,196K,+-0.1%,0.1W,50PPM,0805

R15

802741

RESISTOR,SMR,RES,MF,6.65K,+-0.1%,0.1W,50PPM,0805

R16

783274

RESISTOR,SMR,RES,CERM,10M,+-5%,.125W,400PPM,1206

R17

642329

RESISTOR,SMR,RES,CERM,34.8K,+-1%,.063W,100PPM,0603

R18

1541369

RESISTOR,METAL FILM,150K,+-0.1%,0.1W,25PPM,0805,TAPE

R2

802758

RESISTOR,SMR,RES,VAR,CERM,2K,+-20%,0.25W

R20

802642

RESISTOR,SMR,RES,CERM,2.4M,+-5%,0.1W,400PPM,0805

R21

928754

RESISTOR,SMR,RES,CERM,2.43K,+-1%,0.1W,100PPM,0805

R22

806240

RESISTOR,SMR,RES,CERM,1.2M,+-5%,.125W,200PPM,1206

R3

912493

RESISTOR,SMR,RES,VAR,CERM,100K,+-25%

R4

802766

RESISTOR,SMR,RES,VAR,CERM,50K,+-20%,0.25W

R5 R10 R13

876016

RESISTOR,SMR,RES,CERM,8.87K,+-1%,.125W,100PPM,1206

R6

642303

RESISTOR,SMR,RES,CERM,3.83K,+-1%,.063W,100PPM,0603

R7

643954

RESISTOR,SMR,RES,MF,30.1K,+-0.1%,0.1W,25PPM,0805

R8

929906

RESISTOR,SMR,RES,CERM,36.5K,+-1%,.125W,100PPM,1206

R9

642220

RESISTOR,SMR,RES,CERM,15.4K,+-1%,.063W,100PPM,0603

S1

772301

SWITCH,SWITCH,SLIDE,3P3T,WITH DETENT

U1

802774

IC,OP AMP,LM4250C,BIPOLAR

VR1

929489

BIPOLAR 4041,SMR,IC,V REF,SHUNT,1.2 V,2%,150 PPM,SOT23

W2

682575

RESISTOR,R05A,RES JUMPER,0.02,0.25W

747501

SCREW,SCREW,PH,PO,STL,4-40,.250

431023

SENSING PROBE AND CABLE ASSY

172080

CABLE TIE

615549

CASE,BOTTOM

129882

SCREW,PHP,4-40X3/16

769034

CASE,TOP

735860

COVER,SWITCH

428441

SHOCK ABSORBER

LIMITED WARRANTY & LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under
normal use and service.  The warranty period is one year and begins on the date of shipment.
Parts, product repairs and services are warranted for 90 days. This warranty extends only to
the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to
fuses, disposable batteries or to any product which, in Fluke’s opinion, has been misused,
altered, neglected or damaged by accident or abnormal conditions of operation or handling.
Fluke warrants that software will operate substantially in accordance with its functional
specifications for 90 days and that it has been properly recorded on non-defective media.
Fluke does not warrant that software will be error free or operate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user
customers only but have no authority to extend a greater or different warranty on behalf of
Fluke. Warranty support is available if product is purchased through a Fluke authorized sales
outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice
Buyer for importation costs of repair/replacement parts when product purchased in one
country is submitted for repair in another country.

Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of
charge repair, or replacement of a defective product which is returned to a Fluke authorized
service center within the warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center or send the
product, with a description of the difficulty, postage and insurance prepaid (FOB Destination),
to the nearest Fluke authorized service center.  Fluke assumes no risk for damage in transit.
Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB
Destination).   If Fluke determines that the failure was caused by misuse, alteration, accident
or abnormal condition of operation or handling,  Fluke will provide an estimate of repair costs
and obtain authorization before commencing the work.  Following repair, the product will be
returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and
return transportation charges (FOB Shipping Point).

THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE.  FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER
ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE
OR ANY OTHER THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or
exclusion or limitation of incidental or consequential damages, the limitations and exclusions
of this warranty may not apply to every buyer.   If any provision of this Warranty is held invalid
or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.

Содержание 80T-150U

Страница 1: ...2 F decreasing linearly to 5 4 F at 58 and 302 F 32 to 59 F and 3 6 F from 32 to 212 F 95 to 122 F decreasing linearly to 7 2 F at 58 and 302 F Sensitivity 80T 150U output 1 mV dc C or F Voltage Standoff 60V dc or 30V ac rms 42 4V peak Settling Time 5 5 seconds to settle within 2 for a 50 change RF Field Exposure to an RF field interferes significantly with this module s capacity to accurately mea...

Страница 2: ...e it to a liquid or gas The DMM will display the temperature in degrees Vary the probe angle and pressure when measuring solid surface temperatures The highest stabilized reading will be the most accurate See the following measuring technique Caution The force exerted on the probe tip should not exceed 15 pounds Measuring Technique Here are some suggestions for improving the accuracy of your tempe...

Страница 3: ...elect mV dc range 3 Select the C position of the switch Immerse the probe tip 2 inches into a mercury thermometer monitored 0 C bath and allow 60 seconds for the readings to stabilize 4 Adjust R3 see Figure 4 to obtain the following reading 0 00 0 05 mV dc 0 0 2 mV dc 5 Select the F position of the switch and adjust R2 to obtain the following reading 32 0 0 1 mV dc 32 0 0 4 mV dc 6 Select the C po...

Страница 4: ...ly in accordance with its functional specifications for 90 days and that it has been properly recorded on non defective media Fluke does not warrant that software will be error free or operate without interruption Fluke authorized resellers shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of F...

Отзывы: