80PK-10 

Type K Thermocouple Pipe Clamp Temperature Probe 

Instruction Sheet 

Introduction 

XW

Warning 

To avoid electric shock, do not 

use this probe when voltages 

exceeding 24 V ac or 60 V dc are 

present. The probe is electrically 

connected to the output terminals. 

The 80PK-10 Type K Thermocouple Pipe Clamp 

Temperature Probe (Figure 1) is designed to reliably 

measure the temperature of pipes 1.25-inch (32 mm) 

to 2.5-inch (64 mm) in diameter, at temperatures 

between -29 

°

C and 149 

°

C (-20 

°

F and 300 

°

F), 

while retaining a long service life. 

Plunger

Sensor

V Notch

 

gcw01.eps 

Figure 1. 80PK-10 Pipe Clamp Temperature Probe 

Specifications 

Measurement Range

: -29 

°

C to 149 

°

C (-20 

°

F to 

300 

°

F) 

Accuracy

±

1.9 

°

C (

±

3.4 

°

F). Surface measurement 

accuracy of 

±

1.9 

°

C includes 

±

1.1

°

C variation in alloy 

plus probe contact and shunting uncertainty of 

±

0.8 

°

C. 

Repeatability

±

0.56 

°

C (

±

°

F) 

Output

: @ 25 

°

C (77 

°

F) = 1.00 mV (Reference 

junction at 0 

°

C) 

Seebeck Coefficient

:  

@ 25 

°

C (77 

°

F) = 40.50 

μ

V / 

°

Measurement Time (Time Constant)

0.5 seconds (5 time constants = 1 complete step 

change, i.e. 2.5 seconds)  

Maximum Voltage Rating

: 24 V ac rms, or 

60 V dc 

Sensor Material

: Chromel-Alumel 

Cable Length

: 40 inches (1 meter) 

Cable Insulation

: Material: PVC 

 

Maximum Temperature

: 105 

°

C (220 

°

F) 

 

Minimum Temperature

: -29 

°

C (20 

°

F) 

Conductor

: K Type AWG #24 stranded 

Probe Body

: PBT 

 

Maximum Temperature

: 149 

°

C (300 

°

F) 

 

Minimum Temperature

: -29 

°

C (20 

°

F) 

Connector

: Mini-thermocouple connector with 

7.925 mm (0.312 in) pin spacing 

Media Limitations

: Must be compatible with 

chromel, alumel, and PBT 