Страница 1: ...o 50 C 32 to 122 F S orage Temperature 40 to 60 C 40 to 14 F Relative Humidity 0 to 90 0 to 35 C 0 to 70 35 to 50 C PIN 795203 JULY 1986 1986 John Fluke Mfg Co Inc All rights reserved Litho in U S A Instruction Sheet SOi 410 DC AC Current Probe Batteries QUANTITY TYPE two 1 5V AAA LIF E 600 hrs typical LOW INDICATION Output voltage is between 0 0 and 99 9 mV with the power switch off load impedanc...

Страница 2: ...However for de current measurements the de offset caused by the residual magnetism will contribute to reading errors The residual magnetism can be minimized as follows see Figure 2 1 Use the thumb and forefinger of one hand to grasp the probe in the middle where the two screw heads are exposed 2 With your other hand compress the handles together then release them suddenly The jaws should come toge...

Страница 3: ...ficient warm up time for stable operation 3 Connect the 80i 410 to the Digital Multimeter input and perform the Battery Test procedure given under the heading OPERATION of this instruction sheet 4 Slide the 80i 410 power switch to the ON position 5 Perform the demagnetizing procedure given under the heading OPERATION of this instruction sheet Snap the jaws at least three times to ensure the residu...

Страница 4: ...h a description of the difficulty postage prepaid to a Fluke Service Center Fluke assumes no risk for damage In transit Fluke will at our option repair or replace the defective accessory free of charge However if we determine tt at the failure was caused by misuse alteration accident or abnormal condition of operation you will be billed for the repair The repaired accessory Y lll be returned trans...