Fluke 79559999 Скачать руководство пользователя страница 1

PN 686953
March 1998 Rev.1, 9/01

1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.

T5-600/T5-1000

Electrical Tester

Service Information

Introduction

This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").

 

Safety information

 

Parts and service information

 

Specifications

 

Cleaning procedure

 

Required equipment

 

Performance tests

 

Parts and accessories list

 

Battery replacement procedure

For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.

Safety Information

 Warning

To avoid possible electric shock or personal injury, follow these guidelines:

  Do not use the tester if it is damaged or operating abnormally. Protection

may be impaired.

  Before each use:

  Make sure the battery door is closed and latched.

  Inspect the tester and test leads. Look for cracks, missing plastic, exposed

metal, or damaged insulation. Replace damaged test leads before using the
tester.

  Verify the tester’s operation by measuring a known voltage.

  Replace the batteries as soon as the low battery indicator (

M

) appears.

  Do not use the tester around explosive gas, vapor or dust.

  Do not apply more than the rated voltage, as marked on the tester, between

terminals or between any terminal and earth ground.

  Refer servicing to qualified personnel.

  Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.

  When using the probes, keep your fingers behind the finger guards on the

probes.

  Connect the common test lead before you connect the live test lead.

Disconnect the live test lead first.

  The display will not display hazardous voltages in display hold. The voltage

indicator LED (

) continues to operate.

®

Shop for Fluke products online at:

1.800.561.8187

www.

MyFlukeStore

.ca

Содержание 79559999

Страница 1: ...n may be impaired Before each use Make sure the battery door is closed and latched Inspect the tester and test leads Look for cracks missing plastic exposed metal or damaged insulation Replace damaged...

Страница 2: ...Category III Pollution Degree 2 T5 1000 1000 V rms Overvoltage Category III 600 V RMS OverVoltage Category IV Pollution Degree 2 Maximum Voltage Between Current Fork and Earth Ground T5 600 and T5 100...

Страница 3: ...nt e 1 65 V dc nominal 600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel and lighting systems in large buildings CAT IV protects against transien...

Страница 4: ...gits e 1000 1000 1 1 2 digits Input Characteristics Input Protection Function T5 600 T5 1000 Input impedance nomial K 600 V rms 1000 V rms 1 M 100 pF ac coupled L 600 V rms 1000 V rms 1 M 100 pF e 600...

Страница 5: ...the coil cause the tester to read 10 times more current than the calibrator supplies 1 Put one edge of the butterfly coil in a vice to hold the coil during testing 2 Set the calibrator to the current...

Страница 6: ...ingle conductor magnet wire Separate into two 5 turn coils Use cable ties to hold coils together Connect to Calibrator Mount Banana plugs on wire end qp7f eps Figure 1 Making the Butterfly Coil Shop f...

Страница 7: ...ator ON Calibrator Output T5 1000 Display Limits AC annunciator ON 600 V 60 Hz 589 to 611 1000 V 60 Hz 983 to 1017 Table 3 DC Voltage Tests T5 600 T5 1000 Step Calibrator Output T5 600 Display Limits...

Страница 8: ...tage to the tester s battery terminals 3 Set the calibrator to 1 k Apply this resistance to the tester s probes 4 Slowly decrease the dc voltage just until the tester s low battery indicator M turns o...

Страница 9: ...tion Sheet International 1621978 H5 Belt Holster Accessory TP1 Probe Set Flat Blade Accessory TP4 Probe Set 4 mm Round Accessory Battery Replacement Figure 2 shows how to replace the batteries Observe...

Отзывы: