Precision Process Calibrator
Calibration Adjustment
33
5522A
CALIBRATOR
100 V PK
MAX
HI
INPUT
LO
Press for
VOLTS/mA
Press and Hold
for Switch Test
Switch Test
Reset
LOOP
HART
PWR
VOLTS/mA
SW TEST
7526A
Calibrator
8508A
gxm022.eps
Figure 17. 10 V Read Adjustment Connections
Step 17: 100 V Read Adjustment
Use the test equipment shown in Table 3. Connect the 5522A V source jacks to the
Product isolated jacks. Use the 8508A DMM to monitor the voltage. See Figure 17 for
the connections.
Enter
17
on the PC. Source three calibration points from 5522A:
•
0 V
•
50 V
•
100 V
For each calibration point, adjust the voltage for the maximum accuracy. Use the DMM
indication as feedback. Push ENTER on the PC when stable to go to the next calibration
point.
Calibration is complete. Enter
18
at the PC to exit calibration mode.