Isolated DC Current
The isolated dc current performance test verifies the accuracy of the isolated dc current
source output from the front panel of the Product. Connect 5522A Normal HI to the
Product Isolated Input LO. Connect the Product Isolated input HI to 8508A input A.
Connect 8508A Input LO to 5522A Normal LO. Use the 8508A to measure the 5522A
output as shown in Figure 2. Take the dc current indication from the 8508A as the
applied current. Verify the measurements listed in Table 10. See Figure 11 for
connections.
Table 10. Isolated DC Current
50 mA Input
8508A Reading
Specification
0.000 mA
0.0010 mA
25.000 mA
0.0035 mA
50.000 mA
0.0060 mA
5522A
CALIBRATOR
20 V PK
MAX
HI
INPUT
LO
Press for
VOLTS/mA
Press and Hold
for Switch Test
Switch Test
Reset
LOOP
HART
PWR
VOLTS/mA
SW TEST
7526A
Calibrator
8508A
gxm019.eps
Figure 11. Isolated DC Current Performance Test Connections
Isolated Loop Power
1.
Connect a 1 k
Ω
, 1%, 1 W resistor across Isolated Input HI and LO.
2.
Set Isolated input to mA.
3.
Push
L
. The upper ISO display shows 24 mA LPWR. The indication on the ISO
lower display must be between 26.300 mA to 21.500 mA to pass.
4.
Push
. The indication must be between 21.336 mA to 17.104 mA to pass.
The verification procedure is complete. If any steps did not pass, adjustment or repair is
necessary.