Automatic Pressure Calibrator
Measure Mode
31
As Left
Push
D
(
As Left
). Repeat the procedure from
As Found
.
Calibration results save to the Test Results menu.
Find Task by Tag
Downloaded tasks from DPCTrack2 or other supported software are shown:
1. Use the up and down arrows to scroll through the tasks.
2. Push
B
(
Page Up
) or
C
(
Page Down
) where necessary.
3. Push
D
(
Recall
) to recall the highlighted task.
Bus Polling
Bus Polling goes back into the bus polling screen and scans for connected HART
devices. If no HART device is found after Polling is complete, push
D
(
Restart Bus
Polling
) to restart bus polling to search of the Product BUS for HART devices. The
Product locates the device.
Measure Mode
The measure mode has these functions:
•
mA dc measure
•
mA dc source
•
mA dc simulate
•
V dc measure
•
Switch (continuity)
The lower left of the display shows the operation mode.
Current Measurement
The Product measures 0 mA to 24 mA. Move the cursor to highlight
Measure Current
.
Push
C
to
Source Current
. Push
D
to
Simulate Current
.