Automatic Pressure Calibrator
HART Menu
29
4. Enter the Lower Range Value (PV LRV) necessary.
5. Push
E
to select.
6. Push
C
(Send) to change the LRV.
Device Diagnostics
Select Device Diagnostics to send the self-diagnostics command to the connected
transmitter and view the transmitter Self Test result.
Calibrate (Ad hoc)
Calibrate (Ad Hoc) calibrates a connected HART device without a downloaded task.
The opening screen shows the task type and source/measure configuration for the
device. The transmitter HART configuration is copied into this configuration screen.
Correct as needed for your test:
1. Push
B
(
PV Calibrate
) to change the measure from mA to the PV source value
of the transmitter.
2. Push
C
(
Disable/Enable Loop Power
).
3. Push
D
(
Continue
) to continue to second configuration screen. These choices
are shown:
•
Tolerance of Range
•
Test Strategy
•
Auto Settling Time
•
Transfer Function
These softkeys are active:
A
– Linear/Square Root Selection
B
– Save as Customized
C
– Leak Test
D
– Continue
The next screen starts the As Found Calibration. These softkeys are active:
A
– Abort
B
– Manual Test
D
– Auto Test