718 Series
Users Manual
18
External Fluke Pressure Module
Compatibility
If inappropriate units are selected, the output of Fluke
700P Pressure Modules can cause the Calibrator display
to overflow (OL), or displays values that are too low to be
read. Refer to Table 7 for appropriate unit and range
compatibility.
Table 7. Fluke Pressure Module Compatibility
Pressure Unit
Module Compatibility
psi
Available on all pressure ranges
inH
2
0
All ranges through 3000 psi
cmH
2
0
All ranges through 1000 psi
bar
15 psi and above
mbar
All ranges through 1000 psi
kPa
Available on all pressure ranges
inHg
Available on all pressure ranges
mmHg
All ranges through 1000 psi
kg/cm
2
15 psi and above