This probe parameter refers to the resistance of the control probe at 0
°
C.
Normally this is set for 100.000 ohms.
ALPHA
This probe parameter refers to the average sensitivity of the probe between 0
and 100
°
C. Normally this is set for 0.00385
°
C-1.
Operating Parameters
The operating parameters menu is indicated by,
Operating parameters menu
Press “SET” to enter the menu. The operating parameters menu contains the
cutout reset mode setting.
Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
cutout reset mode parameter
Press “SET” to access the parameter setting. Normally the cutout is set for
manual mode.
cutout set for manual reset
To change to manual reset mode press “UP” and then “SET”.
cutout set for automatic reset
Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine
the operation of the serial interface. These controls only apply to baths fitted
with the serial interface. The parameters in the menu are — baud rate,
sample period, duplex mode, and linefeed.
PAr
CtorSt
Cto=rSt
Cto= Auto
SErIAL