700G Series
Users Manual
8
Function Lock
When set, access to each of the settable parameters
above can be turned “off” to prevent unauthorized
configuration changes. This is done with password
protection in Supervisory mode. Push
to access
Supervisory mode or
to go back to normal operation.
Supervisory Mode
If necessary, each user-configurable parameter can be
edited when you receive the Product. Some parameters
are locked and must be unlocked to configure them. Use
Supervisory mode to do this.
When you are in the Configure menu, and
FUnC LOCK
is
shown, it means that there are locked parameters.
To disable function lock:
1. Push
.
0 PWRD
is then shown.
2. The password “101” is required to unlock
Supervisory mode. Push
to put in the password
entry. Hold
or
down to move faster through the
selections by a factor of 10. When you stop the
counter, push
and
again to move forward or
backward by a factor of 1. The password is factory
set and cannot be changed.
3. Push
.
From this point each parameter can be locked or
unlocked. Push
and
to select
UnLOC
or
LOC
for
each parameter. To move to the next parameter, push
.
You can access, lock, or unlock these functions:
•
Zero function (enable/disable)
•
Set pressure units (enable/disable)
•
Auto shutdown adjustment (enable/disable)
•
Damping settings (enable/disable)
•
Sample rate setting (enable/disable)
•
Tare setting (enable/disable)
•
Custom engineering units (set scale factor)
When a function is locked, it cannot be accessed or
changed from its current condition until you go to
Supervisory Mode and unlock it.
Available Pressure Ranges
Available pressure ranges are listed in the Specifications
section.