SA= 1
Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample
rate to the displayed value.
SA= 60
New sample period
9.12.3
Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du-
plex or half duplex. With full duplex any commands received by the bath via
the serial interface will be immediately echoed or transmitted back to the de-
vice of origin. With half duplex the commands will be executed but not echoed.
The duplex mode parameter is indicated by,
dUPL
Serial duplex mode parameter
Press “SET” to access the mode setting.
dUP=FULL
Current duplex mode setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
dUP=HALF
New duplex mode setting
9.12.4
Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
LF
Serial linefeed parameter
Press “SET” to access the linefeed parameter.
LF= On
Current linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF= OFF
New linefeed setting