13-1
Chapter 13
Troubleshooting
This section contains information on troubleshooting and CE Comments. This
information pertains to a number of bath models and certain specifics may not
pertain to your model.
13.1 Troubleshooting
In the event that the instrument appears to function abnormally, this section may
help to find and solve the problem. Several possible problem conditions are
described along with likely causes and solutions. If a problem arises, please read
this section carefully and attempt to understand and solve the problem. If the
probe seems faulty or the problem cannot otherwise be solved, contact an
Authorized Service Center (see Section 1.3) for assistance. Be sure to have the
instrument model number, serial number, voltage, and problem description
available.
Problem
Causes and Solutions
The heater indicator LED stays red but the
temperature does not increase
The display does not show “Cut-out” nor
displays an incorrect bath temperature, and
the controller otherwise appears to operate
normally. The problem may be insufficient
heating, no heating at all, or too much
cooling.
•
The heater power setting being too low,
especially at higher operating
•
One or more burned out heaters or
blown heater fuses may also cause this
problem. If the heaters seem to be
burned out, contact an Authorized
Service Center (see Section 1.3) for
assistance.
The controller display flashes “Cut-out” and
the heater does not operate
The display flashes “Cut-out” alternately with
the process temperature.
•
If the process temperature displayed
seems grossly in error, consult the
following problem: ‘The display flashes
“Cut-out” and an incorrect process
Содержание 7008
Страница 6: ...7008 Users Guide iv ...
Страница 8: ...7008 Users Guide vi ...
Страница 10: ...7008 Users Guide viii ...
Страница 18: ...7008 Users Guide 2 2 Figure 1 Bath Assembly ...
Страница 24: ...7008 Users Guide 4 4 ...
Страница 45: ...Controller Operation 9 2 Reset Cut out 9 9 3 Figure 7 Controller Operation Flowchart ...
Страница 58: ...7008 Users Guide 9 16 ...
Страница 62: ...7008 Users Guide 10 4 Table 5 Interface Command Summary ...
Страница 70: ...7008 Users Guide 11 4 ...