Controller Operation
9.7 Proportional Band
9
9-9
Table 4. Proportional Band — Fluid Table
Fluid Temperature
Heater
Setting
Proportional
Band
Stability
Water 30.0
°
C Low 0.04
°
C
±
0.0004
°
C
Water 60.0
°
C Low 0.04
°
C
±
0.001
°
C
Eth-Gly 50%
35.0
°
C Low 0.05
°
C
±
0.0005
°
C
Eth-Gly 50%
60.0
°
C Low 0.05
°
C
±
0.001
°
C
Eth-Gly 50%
100.0
°
C High
0.4
°
C
±
0.007
°
C
Oil 35.0
°
C Low 0.1
°
C
±
0.003
°
C
Oil 60.0
°
C Low 0.2
°
C
±
0.002
°
C
Oil 100
°
C Low 0.2
°
C
±
0.003
°
C
The proportional band adjustment may be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.
+
Access heater power in secondary menu
12
Pct
Heater power in percent
Access proportional band
Pb=0.101C
Proportional band setting
To change the proportional band press “UP” or “DOWN”.
Pb=0.060C
New proportional band setting
To accept the new setting and access the cut-out set-point press “SET”. Pressing
“EXIT” will exit the secondary menu ignoring any changes just made to the
proportional band value.
Accept the new proportional band setting
Содержание 7008
Страница 6: ...7008 Users Guide iv ...
Страница 8: ...7008 Users Guide vi ...
Страница 10: ...7008 Users Guide viii ...
Страница 18: ...7008 Users Guide 2 2 Figure 1 Bath Assembly ...
Страница 24: ...7008 Users Guide 4 4 ...
Страница 45: ...Controller Operation 9 2 Reset Cut out 9 9 3 Figure 7 Controller Operation Flowchart ...
Страница 58: ...7008 Users Guide 9 16 ...
Страница 62: ...7008 Users Guide 10 4 Table 5 Interface Command Summary ...
Страница 70: ...7008 Users Guide 11 4 ...