6-1
Chapter 6
Bath Use
READ BEFORE PLACING THE BATH IN SERVICE
The information in this section is for general information only. It is not designed to
be the basis for calibration laboratory procedures. Each laboratory will need to
write their own specific procedures.
6.1 General
Be sure to select the correct fluid for the temperature range of the calibration.
Bath fluids should be selected to operate safely with adequate thermal properties
to meet the application requirements. Also, be aware that some fluids expand
and could overflow the bath if not watched. Refer to General Operation, section
8, for information specific to fluid selection and to the MSDS sheet specific to the
fluid selected. Generally, baths are set to one temperature and used to calibrate
probes only at that single temperature. This means that the type of bath fluid
does not have to change. Additionally, the bath can be left energized reducing
the stress on the system.
The bath generates extreme temperatures. Precautions must be taken to prevent
personal injury or damage to objects. Probes may be extremely hot or cold when
removed from the bath. Cautiously handle probes to prevent personal injury.
Carefully place probes on a heat/cold resistant surface or rack until they are at
room temperature. It is advisable to wipe the probe with a clean soft cloth or
paper towel before inserting it into another bath. This prevents the mixing of
fluids from one bath to another. If the probe has been calibrated in liquid salt,
carefully wash the probe in warm water and dry completely before transferring it
to another fluid. Always be sure that the probe is completely dry before inserting
it into a hot fluid. Some of the high temperature fluids react violently to water or
other liquid mediums. Be aware that cleaning the probe can be dangerous if the
probe has not cooled to room temperature. Additionally, high temperature fluids
may ignite the paper towels if the probe has not been cooled.
For optimum accuracy and stability, allow the bath adequate stabilization time
after reaching the set-point temperature.
Содержание 7008
Страница 6: ...7008 Users Guide iv ...
Страница 8: ...7008 Users Guide vi ...
Страница 10: ...7008 Users Guide viii ...
Страница 18: ...7008 Users Guide 2 2 Figure 1 Bath Assembly ...
Страница 24: ...7008 Users Guide 4 4 ...
Страница 45: ...Controller Operation 9 2 Reset Cut out 9 9 3 Figure 7 Controller Operation Flowchart ...
Страница 58: ...7008 Users Guide 9 16 ...
Страница 62: ...7008 Users Guide 10 4 Table 5 Interface Command Summary ...
Страница 70: ...7008 Users Guide 11 4 ...