To stop a continuous test, push the test button again. The Tester disconnects the test supply, and
holds the last test result on the display.
Note
1.
Fuse/L-N Pre-Test is performed as part of live tests (Touch Current and
Load/Leakage Current Test). The pre-test verifies the fuse and lead continuity
with a low voltage signal across the appliances phase and neutral pins.
Very low power appliances, or appliances with electronically controlled on/off
switches or with an inductance may fail this test. In this case the symbol
will appear.
To enable you to test these appliances you can skip the Fuse/L-N Loop Pre-
test. To perform the test on appliances that fail the Fuse/L-N Loop Pre-test:
•
Release the appropriate function key, push and hold it until your hear the
second indication beep, but release it before the third long beep.
•
To start a continuous test release the appropriate function key, push and
hold it until your hear a third long beep.
2.
The measurement unit blinks for an active measurement (for example,
Ω
, M
Ω
,
mA) during a continuous test.
3.
The IEC-Lead test cannot be done in the continuous test mode.
Stop a Test
Push
to immediately end a test that is in progress, make the Tester safe, and then show the
idle screen. Test results do not show on the display.
How to Save Test Results
To save the result after a test is complete, push
. The actual test result is saved into
memory. The display shows the number that has been assigned to the record, for example:
5
.
For detailed information refer to
Memory
on page 19.
Visual Inspection
Before any test, check the appliance for:
•
condition of the appliance cables, for example, no cuts, cracks or any physical damage to
the outer insulation layer.
•
condition of the plug, cable securely attached, no signs of overheating, and that the
correct value of fuse is fitted.
•
any signs of damage, and that any mains or control switches will physically switch on
and off.
•
any sockets for signs of overheating or physical damage.