Appliance Tester
Power-Up and Warning Display Messages
7
Power-Up and Warning Display Messages
Warning
To prevent possible electrical shock, fire, or personal injury:
•
Read the safety information before you power up the Tester.
•
Connect an approved three-conductor mains power cord to a grounded
power outlet.
•
Do not use a two-conductor mains power cord unless you install a
protective ground wire to the Product ground terminal before you
operate the Product.
•
Make sure that the Product is grounded before use.
The Tester will power up when you connect it to the mains supply.
Disconnect the mains plug to power down the Tester.
At power up the Tester performs a selftest and shows the software version.
After power up, the display indicates the mains supply voltage. This screen is referred to
throughout this manual as the idle screen.
If the Tester detects special conditions at power up, such as a dangerous condition, a warning
message indicates the condition.
Table 7 is a list of the messages that can show when you power up the Tester. The values are
examples and can differ from actual values shown on the display.
Table 7. Display Messages
Display Explanation
Στ
Selftest
1.18
Software version is shown after power on.
230
VAC
Mains supply voltage, idle screen.
U
N-PE
50
VAC
Mains problem, unplug unit! No test possible.
U
L-N
195
VAC
Mains voltage too low. No test possible.
U
L-N
265
VAC
Mains voltage too high. No test possible.
<
48
Hz
Mains frequency too low. No test possible.
>
52
Hz
Mains frequency too high. No test possible.
Memory
full.
75
Memory nearly full (>75 %).
+ number
Tester failure, contact Fluke support.
+ number
Dangerous Tester failure. Unplug the Tester, prevent it from being used, and contact
Fluke for repair.
U
N-PE
71
The Neutral-Earth voltage is dangerously high. Unplug the Tester!
75
The mains polarity is incorrect. Unplug the Tester!
U
N-PE
77
Mains supply earth connection is missing/open circuit. Unplug the Tester!