Calibration Bath
Digital Communication Interface
49
Command Description
Command Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set serial duplex mode:
du[plex]=f[ull]/h[alf]
FULL or
HALF
Set serial duplex mode to
full
du[plex]=f[ull] du=f
Set serial duplex mode to
half
du[plex]=h[alf] du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to
on
lf[eed]=on lf=on
Set serial linefeed mode to
off
lf[eed]=of[f] lf=of
Calibration Menu
Read C0 calibration
parameter
*c0
*c0
b0: 9
b0: 0
Set C0 calibration parameter
to n
*c0=n *c0=0
–999.9
to
999.9
Read CG calibration
parameter
*cg
*cg
bg: 999.99
bg: 156.25
Set CG calibration
parameter to n
*cg=n *cg-156.25
–999.9
to
999.9
Read low set-point limit
value
*tl[ow]
*tl
tl: 999
tl: –80
Set low set-point limit to n
*tl[ow]=n
*tl=-80
–999.9 to
999.9
Read high set-point limit
value
*th[igh]
*th
th: 999
th: 205
Set high set-point limit to n
*th[igh]=n
*th=205
–999.9 to
999.9
Miscellaneous (not on
menus)
Read firmware version
number
*ver[sion] *ver
ver.9999,9.99
ver.2100,3.56
Read structure of all
commands
h[elp]
h
list of commands
Read heater
fn
f1
f1:9
f1:1