Calibration Bath
Controller Operation
43
The display now shows,
Press “SET” to show the current value of Cg. Use the “UP and ”DOWN" keys
until the value displayed matches the Cg value on the Report of Calibration.
Press “SET” to store the value. The display now shows,
H and L
These parameters set the high and low set-point limits of the bath. These
parameters are factory set to the following: High 550
°
C
±
0
°
C, Low 0
°
C
±
0
°
C.
Note
DO NOT change the values of these parameters from the factory set
values. To do so may present danger of the bath overheating and
causing damage or fire.
To adjust, press “SET” to show the current value of H. Use the “UP” and “DOWN”
keys until the value matches the factory settings listed above. Press “SET to
store the value.
The display now shows,
Press “SET” to show the current value of L. Use the “UP” and “DOWN” keys
until the value matches the factory settings listed above. Press “SET” to store
the value.
The display now shows,
Press “EXIT” to return to displaying the temperature.