6050H
User's Guide
42
Calibration Parameters
The operator of the bath controller has access to a number of the bath calibration
constants namely CTO, C0, CG, H, and L. These values are set at the factory
and must not be altered. The correct values are important to the accuracy and
proper and safe operation of the bath. Access to these parameters is available to
the user only so that in the event that the controller’s memory fails the user may
restore these values to the factory settings. The user should have a list of these
constants and their settings with the manual.
W
Caution
DO NOT change the values of the bath calibration constants
from the factory set values. The correct setting of these
parameters is important to the safety and proper operation of
the bath.
The calibration parameters menu is indicated by,
Calibration parameters menu
Press “SET” five times to enter the menu. The display now shows:
CTO
Parameter CTO sets the calibration of the over-temperature cutout. This
parameter is not adjusted through software. It is adjusted with an internal
potentiometer. The following values are set by the factory: 565
°
C
±
5
°
C.
CO and CG
These parameters calibrate the accuracy of the bath set-point. They are
programmed at the factory when the bath is calibrated and the values are
entered onto the Report of Test.
Note
Do not alter the value of these parameters. If improved bath
accuracy is needed, calibrate R0 and ALPHA according to the
procedure given in the
Calibration
section.
To restore these values, press “SET” to display,
Press “SET” to show the current value of C0.
Use the “UP” and “DOWN” keys until the value displayed matches the C0 value
on the Report of Calibration. Press “SET” to store the value.