6050H
User's Guide
32
Press “SET” to accept the new selection and resume displaying the bath
temperature.
Set the new units and resume temperature display
Secondary Menu
Functions which are used less often are accessed within the secondary menu.
The secondary menu is accessed by pressing “SET” and “EXIT”
simultaneously and then releasing. The first function in the secondary menu is
the heater power display (see Figure 6).
Heater Power
The temperature controller controls the temperature of the bath by pulsing the
heater on and off. The total power being applied to the heater is determined by
the duty cycle or the ratio of heater on time to the pulse cycle time. This value
may be estimated by watching the red/green control indicator light or read
directly from the digital display. By knowing the amount of heating to the bath the
user can tell if the bath is heating up to the set-point, cooling down, or controlling
at a constant temperature. Monitoring the percent heater power lets the user
know how stable the bath temperature is. With good control stability the percent
heating power should not fluctuate more than
±
1 % within one minute.
The heater power display is accessed in the secondary menu. Press “SET” and
“EXIT” simultaneously and release. The heater power will be displayed as a
percentage of full power.
+
Access heater power in secondary menu
Heater power in percent
To exit out of the secondary menu press “EXIT”. To continue on to the
proportional band setting function press “SET”.