Press “SET” to accept the new selection and access the set-point value.
Accept selected set-point memory
Set-point Value
The set-point value can be set with a resolution of 0.01
°
C. The user may want to
adjust the set-point slightly to achieve a more precise bath temperature. The set-
point vernier allows the temperature to be adjusted below or above the
set-point by a small amount with very high resolution. Each of the 8 stored set-
points has an associated vernier setting. The vernier is accessed from the set-
point by pressing “SET”. The vernier setting is displayed as a 6 digit number with
five digits after the decimal point. This is a temperature offset in degrees of the
selected units, C or F.
Set-point 4 value in °C
If the set-point value need not be changed then press “EXIT” to resume
displaying the bath temperature. Press “UP” or “DOWN” to adjust the set-point
value.
Increment memory
New set-point value
When the desired set-point value is reached press “SET” to accept the new value
and access the set-point vernier. If “EXIT” is pressed instead then any changes
made to the set-point will be ignored.
Accept new set-point value
Set-point Vernier
The set-point value can only be set with a resolution of 0.01
°
C. The user may
want to adjust the set-point slightly to achieve a precise bath temperature. The
set-point vernier allows one to adjust the temperature below or above the set-
point by a small amount with very high resolution. Each of the 8 stored set-points
has an associated vernier setting. The vernier is accessed from the set-point by
pressing “SET”. The vernier setting is displayed as a 6 digit number with five
digits after the decimal point. This is a temperature offset in degrees of the
selected units, C or F.
Current vernier value in °C
To adjust the vernier press “UP” or “DOWN”. Unlike most functions the vernier
setting has immediate effect as the vernier is adjusted. “SET” need not be
pressed. This allows one to continually adjust the bath temperature with the
vernier as it is displayed.