Calibration Bath
Controller Operation
27
The controller may be operated in temperature units of degrees Celsius or
Fahrenheit. The controller is operated and programmed from the front control
panel using the four key switches and digital LED display. The controller may
also be optionally equipped with a serial RS-232 or IEEE-488 GPIB digital
interface for remote operation. Operation of the controller using the front control
panel is discussed following in the
Controller Operation
section. Operation using
the digital interfaces is discussed in the
Digital Communications Interface
section.
When the controller is set to a new set-point the bath heats or cools to the new
temperature. Once the new temperature is reached the bath usually takes 10–15
minutes for the temperature to settle and stabilize. There may be a small
overshoot or undershoot of about 0.5
°
C.
Controller Operation
This section discusses in detail how to operate the bath temperature controller
using the front control panel. Using the front panel key switches and LED display
the user may monitor the bath temperature, set the temperature set-point in
degrees C or F, monitor the heater output power, adjust the controller
proportional band, set the cut-out set-point, and program the probe calibration
parameters, operating parameters, serial and IEEE-488 interface configuration,
and controller calibration parameters. Controller operations are summarized in
the flow chart in Figure 6.
Bath Temperature
The digital LED display on the front panel allows direct viewing of the actual bath
temperature. This temperature value is what is normally shown on the display.
The units, C or F, of the temperature value are displayed at the right. For
example,
Bath temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
Reset Cutout
If the over-temperature cutout has been triggered, then the temperature display
alternately flashes,
Indicates cutout condition
The message continues to flash until the temperature is reduced and the cutout
is reset.
The cutout has two modes — automatic reset and manual reset. The mode
determines how the cutout is reset which allows the bath to heat up again. When
in automatic mode, the cutout resets itself as soon as the temperature is lowered
below the cutout set-point. With manual reset mode the cutout must be reset by
the operator after the temperature falls below the set-point.