Calibration Bath
Installation
15
Power
With the bath power switch off (see Figure 3), plug the bath into an AC mains
outlet of the appropriate voltage, frequency, and current capacity (see the
Specifications section for power details). A separate ground connection is
provided and required to permanently connect the instrument to earth ground for
added operator safety.
Be sure the stirring motor power cord is plugged into the “STIRRER” socket at
the back of the bath (see Figure 5).
Tipping Prevention Brackets Installation
W
Warning
Fluke Calibration recommends installing the bath on a heat-
resistant non-flammable surface.
In order to create the safest possible conditions under use, your bath comes
equipped with Tipping Prevention Brackets. The installation of the brackets is
required for compliance with the International Safety Standard IEC 61010-1,
Section 7.3, Stability, which applies to the stability of the bath under normal
operating conditions.
Installation on a Concrete Floor
Using a concrete drill and concrete drill bit, drill three 1/4 inch x 1 1/4 inch deep
(approximately 6.5 mm x 32 mm) holes in the concrete floor using the bracket to
mark the hole placement. Drop the flare anchor bolt into the hole. Tightening the
screw expands the anchor in the drilled hole and secures the bracket. Ensure
that the bracket is installed in such a way as to ensure the bath will have a
minimum of 6 inches of clearance for air circulation (see Figure 2). Screw the
brackets securely to the floor.
Installation of The Bath
Slide the back casters of the bath completely into the bracket caster slots (see
Figure 2). Lock the front casters of the bath. Check that the bath is securely
locked into the brackets by gently pushing on the bath. Proceed to fill the bath
with the applicable bath fluid after reading the entire User's Guide. If you have
any questions concerning installation of the tipping prevention brack
ets, please
contact an Authorized Service Center (see the
How to Contact Fluke Calibration
section).