6050H
User's Guide
12
Press “SET” to select this memory and access the set-point value.
Access set-point value
Current value of set-point 1, 25.00
°
C
Press “UP” or “DOWN” to change the set-point value.
Increment display
New set-point value
Press “SET” to accept the new value and display the vernier value. The bath
begins heating or cooling to the new set-point.
Store new set-point, access vernier
Current vernier value
Press “EXIT” and the bath temperature will be displayed again.
Return to the temperature display
Bath temperature display
The bath heats or cools until it reaches the new set-point temperature. Set the
heater switch to position “MED” or “HIGH” to allow the bath to more quickly reach
a higher temperature. The higher heater settings may be necessary to reach
higher temperatures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set to
prevent this from happening. See the
Cutout
section.
If operating the bath below 45
°
C set the COOLING power switch to ON. The
cooling temperature may require adjustment to provide the proper amount of
cooling. See the
Cooling
section.
To obtain optimum control stability adjust the proportional band as discussed in
the
Proportional Band
section.