Calibration Bath
Specifications and Environmental Conditions
9
Specifications and Environmental Conditions
Specifications
6050H
Range 180
°
C to 550
°
C †
Stability
±
0.002
°
C at 200
°
C (salt)
±
0.004
°
C at 300
°
C (salt)
±
0.008
°
C at 550
°
C (salt)
Uniformity
±
0.005
°
C at 200
°
C (salt)
±
0.020
°
C at 550
°
C (salt)
Temperature Setting
Digital display with push-button data entry
Set-Point Resolution
0.01
°
C; high-resolution mode, 0.00018
°
C
Display Temperature Resolution
0.01
°
C
Digital Setting Accuracy
±
1
°
C
Digital Setting Repeatability
±
0.02
°
C
Heaters 400/1200/2000
Watts
Access Opening
5" x 10" (127 x 254 mm)
Depth
12" (305 mm)
Wetted Parts
304 stainless steel
Power
230 VAC (
±
10 %), 50/60 Hz, 10 A, 2200 W
Volume
7.1 gallons (27 liters), requires 112 lb. of bath salt
Weight
180 lb. (82 kg)
Size
28.5" H x 20.4" W x 24.5" D
(724 x 518 x 622 mm)
Automation Package
Interface-it software and RS-232 computer
interface are available for setting bath
temperature via remote computer. For IEEE-488,
add the
2001-IEEE to the automation package.
Safety
OVERVOLTAGE (Installation) CATEGORY II,
Pollution Degree 2 per IEC61010-1
Содержание 6050H