5500A
Service Manual
1-6
1-4. General
Specifications
Warmup Time
Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time
Less than 5 seconds for all functions and ranges except as noted.
Standard Interfaces
IEEE-488 (GPIB), RS-232, 5725A Amplifier
Temperature
Performance
•
Operating: 0 °C to 50 °C
•
Calibration (tcal): 15 °C to 35 °C
•
Storage: -20 °C to 70 °C
Temperature Coefficient
Temperature Coefficient for temperatures outside tcal +
5
°C is 0.1X/°C of
the 90-day specification (or 1-year, as applicable) per °C.
Relative Humidity
[1]
•
Operating: <80% to 30 °C, <70% to 40 °C, <40% to 50 °C
•
Storage: <95%, non-condensing
Altitude
•
Operating: 3,050 m (10,000 ft) maximum
•
Non-operating: 12,200 m (40,000 ft) maximum
Safety
Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation
20 V
EMC
Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in
areas with Electromagnetic fields of 1 to 3 V/m, resistance outputs have a
floor adder of 0.508
Ω
. Performance not specified above 3 V/m. This
instrument may be susceptible to electro-static discharge (ESD) from
direct contact to the binding posts. Good static aware practices should be
followed when handling this and other pieces of electronic equipment.
Line Power
•
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
•
Line Frequency: 47 Hz to 63 Hz
•
Line Voltage Variation: ±10% about line voltage setting
Power Consumption
5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
Dimensions
5500A Calibrator:
•
Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
•
Width: 43.2 cm (17 inches), standard rack width
•
Depth: 47.3 cm (18.6 inches) overall 5725A Amplifier:
•
Height, 13.3 cm (5.25 inches), standard rack increment, plus 1.5 cm
(0.6 inch) for feet on bottom of unit;
•
Width, 43.2 cm (17 inches), standard rack width
•
Depth, 63.0 cm (24.8 inches) overall.
Weight (without options)
5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 pounds)
Absolute Uncertainty
Definition
The 5500A specifications include stability, temperature coefficient,
linearity, line and load regulation, and the traceability of the external
standards used for calibration. You do not need to add anything to
determine the total specification of the 55
0
0A for the temperature range
indicated.
Specification
Confidence Interval
99%
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be
required.
Содержание 5500A
Страница 12: ...5500A Service Manual x...
Страница 16: ...5500A Service Manual 1 2...
Страница 46: ...5500A Service Manual 1 32...
Страница 48: ...5500A Service Manual 2 2...
Страница 98: ...5500A Service Manual 4 2...
Страница 101: ...Maintenance Access Procedures 4 4 5 om016f eps Figure 4 1 Exploded View of Rear Panel Assemblies...
Страница 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Страница 116: ...5500A Service Manual 5 2...
Страница 120: ...5500A Service Manual 5 6 MP3 MP14 4X MP26 MP2 H13 X16 5500A Final Assembly 5 of 6 om018f eps Figure 5 1 Chassis Assembly...
Страница 127: ...6 1 Chapter 6 Oscilloscope Calibration Options Option 5500A SC600 see page 6 3 Option 5500A SC300 see page 6 65...
Страница 128: ...5500A Service Manual 6 2...
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 268: ......
Страница 269: ......
Страница 270: ......
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 277: ......
Страница 278: ......
Страница 279: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ...DS2E SL2 DC5V RELAY...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......