5500A
Service Manual
6-28
13.
The Calibrator Mainframe will prompt that the calibration is complete. Press the
OPTIONS, then STORE CONSTS blue softkeys to store the new calibration
constants.
6-43. Verification
All of the Oscilloscope Calibration functions should be verified at least once per year, or
each time the SC600 is calibrated. The verification procedures in this section provide
traceable results; however the factory uses different procedures and instruments of
higher precision than those described here. The procedures in this manual have been
developed to provide users the ability to verify the SC600 at their own site if they are
required to do so. Fluke strongly recommends that, if possible, you return your unit to
Fluke for calibration and verification.
All equipment specified for SC600 verification must be calibrated, certified traceable if
traceability is to be maintained, and operating within their normal specified operating
environment. It is also important to ensure that the equipment has had sufficient time to
warm up prior to its use. Refer to each equipment’s operating manual for details.
Before you begin verification, you may wish to review all of the procedures in advance
to ensure you have the resources to complete them.
All of the SC600 functions are listed in Table 6-18, with the verification technique
indicated.
Table 6-18. Verification Methods for SC600 Functions
Function
Verification Method
DC Voltage
Procedure provided in this manual.
AC Voltage amplitude
Procedure provided in this manual.
AC Voltage frequency
Procedure provided in this manual.
Edge amplitude
Procedure provided in this manual.
Edge frequency, duty
cycle, rise time
Procedure provided in this manual.
Tunnel Diode Pulser
amplitude
Procedure provided in this manual. See “Voltage and Edge Calibration
and Verification” for details.
Leveled sine wave
amplitude, frequency,
harmonics, and flatness
Procedures provided in this manual.
Time marker period
Procedure provided in this manual.
Wave generator
amplitude
Procedure provided in this manual.
Pulse width, period
Procedure provided in this manual.
MeasZ resistance,
capacitance
Procedure provided in this manual.
Overload functionality
Procedure provided in this manual.
Содержание 5500A
Страница 12: ...5500A Service Manual x...
Страница 16: ...5500A Service Manual 1 2...
Страница 46: ...5500A Service Manual 1 32...
Страница 48: ...5500A Service Manual 2 2...
Страница 98: ...5500A Service Manual 4 2...
Страница 101: ...Maintenance Access Procedures 4 4 5 om016f eps Figure 4 1 Exploded View of Rear Panel Assemblies...
Страница 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Страница 116: ...5500A Service Manual 5 2...
Страница 120: ...5500A Service Manual 5 6 MP3 MP14 4X MP26 MP2 H13 X16 5500A Final Assembly 5 of 6 om018f eps Figure 5 1 Chassis Assembly...
Страница 127: ...6 1 Chapter 6 Oscilloscope Calibration Options Option 5500A SC600 see page 6 3 Option 5500A SC300 see page 6 65...
Страница 128: ...5500A Service Manual 6 2...
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......
Страница 264: ......
Страница 265: ......
Страница 266: ......
Страница 267: ......
Страница 268: ......
Страница 269: ......
Страница 270: ......
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ......
Страница 277: ......
Страница 278: ......
Страница 279: ......
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ...DS2E SL2 DC5V RELAY...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......