5522A
Operators Manual
9-30
TZ1MOHM
Sets the impedance to 1 M in Meas Z mode.
TZCAP
Sets the impedance to cap in Meas Z mode.
TOLDC
Sets the instrument to DC in Overload mode.
TOLAC
Sets the impedance to AC in Overload mode.
Example:
RANGE TP20DB
Edge Function Commands
TDPULSE
(IEEE-488, RS-232, Sequential)
Turns tunnel diode pulse drive on/off in EDGE mode.
Parameters:
ON
(or non-zero) or
OFF
(or zero)
Example:
TDPULSE ON
Returns the tunnel diode pulse drive setting in EDGE mode.
Parameters: None
Response:
1 if ON, 0 if OFF.
Marker Function Commands
TMWAVE
(IEEE-488, RS-232, Sequential)
Selects the waveform for MARKER mode.
Parameters:
SINE
Sine wave (2 ns to 15 ns)
SPIKE
Triangular/sawtooth pulse (15 ns to 5s)
SQUARE
Square wave (50% duty cycle) (4 ns to 5s)
SQ20PCT
Square wave (20% duty cycle) (85 ns to 5s)
Example:
TMWAVE SPIKE
TMWAVE?
(IEEE-488, RS-232, Sequential)
Returns the MARKER mode waveform setting.
Parameters: None
Response: <character> (Returns
SINE, SPIKE, SQUARE
, or
SQ20PCT
.)
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...