Remote Operations
How to Use Commands
5
5-27
Table 5-4. IEEE-488 Interface Messages (Received) (cont.)
Mnemonic Name
Function
MTA
My Talk Address
Addresses a specific device on the bus as a talker. The
controller sends MTA automatically whenever it directs a
device-dependent or common query to a specific
instrument.
REN
Remote Enable
Transfer remote/local control of the Calibrator. (See Table
5-1)
RFD
Ready For Data
Sets the handshake signal line NRFD low.
SDC
Selected Device Clear Does the same thing as DCL, but only if the Calibrator is
currently addressed as a listener.
SPD
Serial Poll Disable
Cancels the effect of a Serial Poll Enable.
SPE
Serial Poll Enable
After the Calibrator receives this message, it sends the
Status Byte the next it is addressed as a listener, no matter
what the command is.
UNL
Unlisten
“Unaddresses” a specific device on the bus as a listener.
The controller sends UNL automatically after the device has
successfully received a device-dependent or common
command.
UNT
Untalk
“Unaddresses” a specific device on the bus as a listener.
The controller sends UNL automatically after the device has
successfully received a device-dependent or common
query.
Table 5-5. IEEE-488 Interface Messages (Sent)
Mnemonic Name
Function
END
End
A message that occurs when the Calibrator asserts the EOI
control line. The Calibrator asserts EOI while it transmits the
ASCII character LF for its termination sequence or
terminator.
DAC
Data Accepted
Set the handshake signal line NDAC low.
DAV
Data Valid
Asserts the handshake signal line DAV.
RFD
Ready for Data
Sets the handshake line NRFD low.
SRQ
Service Request
A control line that any device on the bus can assert to
indicate that it requires attention. Refer to “Checking
Calibrator Status” for details.
STB
Status Byte
The status byte is what the Calibrator sends when it
responds to a serial poll (interface message SPE).
Compound Commands
A compound command is two or more commands in a single command line. For
example, the following two commands could be entered individually,
OUT 1 V, 60 HZ
OPER
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...