39
8 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set decimal place
dp=n
dp=1
1 or 2
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=n
sa=0
0 to 999
Set serial duplex mode:
du[plex]=f[ull]/h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu
Read firmware version number
*ver[sion]
*ver
ver.9999,9.99
ver.9141,1.21
Read structure of all commands
h[elp]
h
list of commands
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Interface Commands continued