Insulation Multimeters
Basic Maintenance
7
Testing the Fuse
XW
Warning
To avoid electrical shock or injury, remove the test leads
and any input signals from the Meter before replacing the
fuse.
Test the fuse as described below and illustrated in Figure 1. If the fuse is defective,
see
Replacing the Batteries and/or Fuse
later in this manual.
1.
Insert a test probe in the
J
input terminal.
2.
Turn the rotary switch to
N
and verify the Meter is in Auto Range.
3.
Insert the probe in the mA
input terminal. If the display reading is
0L
, the fuse is
defective; replace the fuse.
TEST
MIN MAX
HOLD
LO
440 mA
OK
ebp01f.eps
Figure 1. Testing the Fuse