1587/1577
Calibration
4
DC Voltage Measurement
Range Resolution
Accuracy 1587
[1]
±
(% of Rdg + Digits)
Accuracy 1577
[1]
±
(% of Rdg + Digits)
6.000 V dc
0.001 V
±
(0.09 % + 2)
±
(0.2 % + 2)
60.00 V dc
0.01 V
±
(0.09 % + 2)
±
(0.2 % + 2)
600.0 V dc
0.1 V
±
(0.09 % + 2)
±
(0.2 % + 2)
1000 V dc
1 V
±
(0.09 % + 2)
±
(0.2 % + 2)
[1]
Accuracies apply to
±
100% of range.
Input Impedance ................................................. 10 M
Ω
(nominal), <100 pF
Normal Mode Rejection Ratio............................. >60 dB @ 50 Hz or 60 Hz
Common Mode Rejection Ratio.......................... >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)
Overload Protection ............................................ 1000 V rms or dc
DC Millivolts Measurement
Range Resolution
Accuracy 1587
±
(% of Rdg + Digits)
Accuracy 1577
±
(% of Rdg + Digits)
600.0 mV dc
0.1 mV
±
(0.1 % + 1)
±
(0.2 % + 1)
DC and AC Current Measurement
Range Resolution
Accuracy 1587
±
(% of Rdg+Digits)
Accuracy 1577
±
(% of Rdg+Digits)
Burden Voltage
(Typical)
400 mA
0.1 mA
±
(1.5 % + 2)
[1]
±
(2 % + 2)
[1]
AC
45 Hz to
1000 Hz
60 mA
0.01 mA
±
(1.5 % + 2)
[1]
±
(2 % + 2)
[1]
2 mV/mA
400 mA
0.1 mA
±
(0.2 % + 2)
±
(1.0 % + 2)
DC
60 mA
0.01 mA
±
(0.2 % + 2)
±
(1.0 % + 2)
2 mV/mA
[1]
1 kHz bandwidth.
Overload ............................................................ 600 mA for 2 minutes maximum
Overload Protection ............................................ 440 mA, 1000 V, FAST fuse
AC Conversion.................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave
input. Conversions are true-rms responding and specified from 5
% to 100 % of range. Input signal crest factor can be up to 3 up to
300 mA, decreasing linearly to crest factor <= 1.5 at 600 mA. For
non-sinusoidal waveforms add +(2 % r 2 % FS) typical,
for a crest factor up to 3.
Ohms Measurement
Range Resolution
Accuracy 1587
[1]
+(% of Rdg+Digits)
Accuracy 1577
[1]
+(% of Rdg+Digits)
600.0
Ω
0.1
Ω
6.000 k
Ω
0.001
k
Ω
60.00 k
Ω
0.01
k
Ω
600.0 k
Ω
0.1
K
Ω
6.000 M
Ω
0.001
M
Ω
±
(0.9 % + 2)
±
(1.2 % + 2)
50.0 M
Ω
0.01
M
Ω
±
(1.5 % + 3)
±
(2.0 % + 3)
[1]
Accuracies apply from 0 to 100% of range.
Overload Protection............................................ 1000 V rms or dc
Open Circuit Test Voltage................................... <8.0 V dc
Short Circuit Current ........................................... <1.1 mA
Diode Test (1587 Only)
Diode Test Indication .......................................... Display voltage drop: 0.6 V at 1.0 mA nominal test current:
Accuracy .............................................................
±
(2 % + 3)