Insulation Multimeters
Electrical Specifications
3
Electrical Specifications
AC Voltage Measurement
1587 Accuracy
Range Resolution
50 Hz to 60 Hz
±
(% of Rdg + Digits)
60 Hz to 5000 Hz
±
(% of Rdg + Digits)
600.0 mV
0.1 mV
±
(1 % + 3)
±
(2 % + 3)
6.000 V
0.001 V
±
(1 % + 3)
±
(2 % + 3)
60.00 V
0.01 V
±
(1 % + 3)
±
(2 % + 3)
600.0 V
0.1 V
±
(1 % + 3)
±
(2 % + 3)
[1]
1000 V
1 V
±
(2 % + 3)
±
(2 % + 3)
[1]
[1]
1 kHz bandwidth.
1587 Low-pass Filter Voltage
Range Resolution
50 Hz to 60 Hz
±
(% of Rdg + Digits)
60 Hz to 400 Hz
±
(% of Rdg + Digits)
600.0 mV
0.1 mV
±
(1 % + 3)
+(2 % + 3)
-(6 % - 3)
6.000 V
0.001 V
±
(1 % + 3)
+(2 % + 3)
-(6 % - 3)
60.00 V
0.01 V
±
(1 % + 3)
+(2 % + 3)
-(6 % - 3)
600.0 V
0.1 V
±
(1 % + 3)
+(2 % + 3)
-(6 % - 3)
1000 V
1 V
±
(2 % + 3)
+(2 % + 3)
-(6 % - 3)
1577 Accuracy
Range Resolution
50 Hz to 60 Hz
±
(% of Rdg + Digits)
600.0 mV
0.1 mV
±
(2 % + 3)
6.000 V
0.001 V
±
(2 % + 3)
60.00 V
0.01 V
±
(2 % + 3)
600.0 V
0.1 V
±
(2 % + 3)
1000 V
1 V
±
(2 % + 3)
AC Conversion.................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave
input. Conversions are true-rms responding and specified from 5
% to 100 % of range. Input signal crest factor can be up to 3 at up
to 500 V, decreasing linearly to crest factor <= 1.5 at 1000 V. For
non-sinusoidal waveforms add
±
(2 % r 2 % FS) typical,
for a crest factor up to 3.
Input Impedance ................................................. 10 M
Ω
(nominal), <100 pF, ac-coupled
Common Mode Rejection Ratio
(1 k
Ω
unbalanced) .......................................... >60 dB at dc, 50 or 60 Hz
Overload Protection ............................................ 1000 V rms or dc, 10
7
V Hz Max