RUSKA 7615
Users Manual
6-4
2.
When the Measured pressure stabilizes, use the 7615’s numeric keypad and OK to
enter the actual pressure applied by the calibration standard.
Do not enter the
Measured pressure reported by the Calibrator
. If necessary, use the CLEAR key to
correct a mistake in the edit field. If the actual pressure applied is acceptable, the
Calibration, Step 3, screen will appear.
Note
If the actual pressure is outside of the tolerance for the requested mid-point
pressure, Error—222 Data out of range will occur. Acknowledge this error
by selecting OK, then re-enter the actual pressure, repeating Step 2,
number1, if necessary.
Step 3
1.
To begin Step 3, use the calibration standard to Apply the high-point pressure of the
range of the sensor being calibrated requested by the 7615. As pressure is admitted
into the Test Port, the Measured pressure on the Calibrator’s screen will change
accordingly.
2.
When the Measured pressure stabilizes, use the Calibrator’s numeric keypad and OK
to enter the actual pressure reported by the calibration standard.
Do not enter the
Measured pressure reported by the Calibrator
. If necessary, use the Clear key to
correct a mistake in the edit field. If the actual pressure applied is acceptable, the
Calibration, Step 4, screen will appear.
Note
If the actual pressure applied is outside of the tolerance for the requested
high-point pressure, Error—222 Data out of range will occur.
Acknowledge this error by selecting OK, then re-enter the actual pressure,
repeating Step 3, number 1, if necessary.
Step 4
1.
To begin Step 4, use the calibration standard to again
Apply
the mid-point pressure
of the range of the sensor being calibrated requested by the Calibrator. As pressure is
admitted into the Test Port, the
Measured pressure
on the Calibrator’s screen will
change accordingly.
2.
When the Measured pressure stabilizes, use the Calibrator’s numeric keypad and
OK
to enter the
actual pressure
reported by the calibration standard.
Do not enter the
Measured pressure reported by the Calibrator.
If necessary, use the
CLEAR
key to
correct a mistake in the edit field. If the
actual pressure applied
is acceptable, the
Calibration complete
screen will appear.
Storing the Coefficients
Note
In addition to saving the calibration coefficients to the Calibrator’s memory,
the user is advised to separately record the calibration coefficients and
store this “backup” in a safe place.
Step 5
Calibration is complete. To exit the calibration procedure without storing the calibration
coefficients in memory, press
Cancel
. To store the calibration coefficients, select
OK
.
Содержание RUSKA 7615
Страница 8: ...RUSKA 7615 Users Manual vi...
Страница 10: ...RUSKA 7615 Users Manual viii...
Страница 16: ...RUSKA 7615 Users Manual 1 6...
Страница 22: ...RUSKA 7615 Users Manual 2 6...
Страница 40: ...RUSKA 7615 Users Manual 4 12...
Страница 48: ...RUSKA 7615 Users Manual 5 8...
Страница 62: ...RUSKA 7615 Users Manual A 4...