RUSKA 7615
Users Manual
3-4
4.
Another method to fill or purge the device under test is to use the system’s “purge
fill” option. To use this option, from the front panel main menu press MENU, TEST,
PURGE and FILL PUMP. Enter the number of Fill Cycles (Charge/discharge) you
would like to run. Break loose the connection at the top of manifold holding device
under test and then press ENTER to start the operation. This will fill and drive out the
air from the loose connection. Stop the system as soon as clear fluid starts coming out
of the loose connection and then tighten the connection.
Tutorial
At this point, the Calibrator should be in the power-up state and the hydraulic/pneumatic
connections ready. The Calibrator should be displaying a screen similar to the one shown
below. (If the bottom line of the display is not showing these options press [
F6
] if an
error message is displayed, then press
Prev
).
MEASURE
27.16
PSI
Set Point
0.00
CONTROL
VENT
STEP
STEP
MENU
MEASURE
ABSOLUTE
This is the Main Menu. It is at the top level of the menu tree and all descriptions in this
manual start from this point. The top center of the display shows the current pressure
(27.16 psi). The upper left corner shows the current mode of the Calibrator (Measure or
Control). The right side of the screen shows a bar graph displaying the current pressure
relative to a user-configurable full-scale value. Below the pressure is the pressure control
Set Point with a numeric scratchpad for entering new Set Point value. The bottom line of
the screen displays the current assignments of the function keys [
F1
] through [
F6
] that
are located below the display.
1.
The first thing we will do is change the pressure units. The units are changed from
the screen MENU/SETUP/UNITS. This means from the Main Menu press Menu
(the F6 key). This will display the MENU screen.
MEASURE
27.16
PSI
Set Point
0.00
Tare
Setup
Cal.
Program
Test
Disp
ABSOLUTE
2.
Now press Setup (the F2 key). This will display the MENU/SETUP screen.
MEASURE
27.16
PSI
Set Point
0.00
Limits
User
Units
Remote
System
ABSOLUTE
Содержание RUSKA 7615
Страница 8: ...RUSKA 7615 Users Manual vi...
Страница 10: ...RUSKA 7615 Users Manual viii...
Страница 16: ...RUSKA 7615 Users Manual 1 6...
Страница 22: ...RUSKA 7615 Users Manual 2 6...
Страница 40: ...RUSKA 7615 Users Manual 4 12...
Страница 48: ...RUSKA 7615 Users Manual 5 8...
Страница 62: ...RUSKA 7615 Users Manual A 4...